Matrix adjustment, check and calibration
4 - 1
ba75651e03
08/2009
4
Matrix adjustment, check and calibration
4.1
General information
Calibration free
operation
The AmmoLyt
®
Plus
700 IQ ammonium sensor is immediately ready to
measure after being equipped with electrodes. For precise
measurements, it is only necessary to adjust the electrodes to the
sample matrix ("matrix adjustment"). In the recommended application
(see section 1.3 R
), the measuring
characteristics of the electrodes remain stable for their entire service
life. Thus, calibration is no longer required.
Possible changes of the sample matrix can be determined by
occasional comparison measurements (e.g. photometer) and
compensated for by a new matrix adjustment as necessary. The sensor
does not have to be taken out of the sample for this. At the same time,
a matrix adjustment provides information on the state of the electrodes.
Steps
Fig. 4-1
Schematic diagram
Procedures for special cases
Calibration free operation
Yes if a new or different electrode is
commissioned
Содержание AmmoLyt Plus 700 IQ
Страница 1: ...ba75651e03 08 2009 AmmoLyt Plus 700 IQ Operating manual IQ SENSOR NET Modular sensor for ammonium...
Страница 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Страница 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Страница 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Страница 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Страница 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...