Check and calibration in standard solutions
4.3.1
General information on checking and calibrating
In addition to the comparison measurement or matrix adjustment under
real measurement conditions, it is possible to check the function of the
entire sensor in standard solutions. If necessary, you can also take
over in the sensor the slope and potential level of the individual
combination electrodes via an exact calibration in standard solutions.
The slope will be retained with a future matrix adjustment if it is carried
out without zeroing the electrode.
Caution
The long term stable VARiON
®
Plus
electrodes are calibration-free
in the recommended application. A check in standard solutions is
only of significance if all basic conditions (cleanness,
conditioning etc.) are strictly kept. Calibration can cause major
measurement errors if the basic conditions are not sufficiently
kept. After calibrating, a matrix adjustment is normally required
additionally (always with the recommended application).
Standard solutions do not correspond to any real test sample!
When does a check or
calibration make sense?
A check or calibration can be useful in the following special cases:
z
If the measured values do not appear to be correct even after a
careful matrix adjustment and if you suspect the electrode slopes to
have changed
z
If a new application is to be started, the sample composition of which
considerably deviates from that of the recommended application
(see section 1.3 R
z
Routinely within the framework of the company quality assurance.
Содержание AmmoLyt Plus 700 IQ
Страница 1: ...ba75651e03 08 2009 AmmoLyt Plus 700 IQ Operating manual IQ SENSOR NET Modular sensor for ammonium...
Страница 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Страница 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Страница 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Страница 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Страница 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...