Matrix adjustment, check and calibration
AmmoLyt
®
Plus
700 IQ
4 - 4
ba75651e03
08/2009
Practical instructions
z
The ammonium concentration has to be determined immediately
after taking the sample as the ammonium content changes very
quickly due to the micro organisms that are present. It is best to take
the sample using a syringe filter for transport to the laboratory or to
stabilize it otherwise. When adding stabilizing solutions, the dilution
factor has to be taken into account.
z
While determining the reference concentrations in the lab (step 2),
you can use the sensor for control purposes again by switching the
maintenance condition off and thus releasing the linked outputs. The
sensor continues to use the data of the previous matrix adjustment
(or calibration). The reference voltages determined in step 1 will not
be lost. They remain stored until step 3 of the matrix adjustment is
completed. They do not have to be noted and entered again.
z
Use the
Online help
if you are unsure of anything during the matrix
adjustment.
z
The AmmoLyt
®
Plus
700 IQ enables to use the IQ-LabLink procedure
for the simplified data transfer by means of a USB memory device
between the IQ S
ENSOR
N
ET
and a laboratory photometer (e.g.
photoLab
®
6000 / spectroFlex - series) (see section 4.2.3).
Содержание AmmoLyt Plus 700 IQ
Страница 1: ...ba75651e03 08 2009 AmmoLyt Plus 700 IQ Operating manual IQ SENSOR NET Modular sensor for ammonium...
Страница 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Страница 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Страница 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Страница 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Страница 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...