17
Technical Manual
Scale Filtering Settings
Sub-block 4.x
Parameter
Value
Description
P4.0
Filter Preset
1
=
Light
2 = Medium <
3 = Heavy
4 = Animal 1
Adjusts filter parameters
P4.1 to P4.4
to a preset
value. Use this to quickly find a starting point for
scale filtering. The User can then fine tune the
filtering if required using the parameters below.
P4.1
Filter
Frequency
0.5 Hz
1 Hz
2 Hz <
3 Hz
Frequency in Hertz of the front end digital filter. A
lower frequency will make the scale more
immune to vibrations, but will also slow down the
response time of the display.
P4.2
A/D Averaging
5
10
50 <
75
100
150
200
Selects the number of A/D conversions that are
averaged to obtain a displayed reading. A higher
number gives a more accurate display by
reducing noisy readings, but slows down the
settling rate of the display.
P4.3
A/D Averaging
Cut-Out
Threshold
2d <
4d
8d
12d
14d
18d
Sets the weight change threshold in displayed
divisions where the A/D averaging is suspended.
This will make the display more responsive to
weight changes above the cut-out threshold.
P4.4
A/D Averaging
Cut-Out
Sensitivity
2
5 <
8
10
12
15
Specifies the number of consecutive A/D
samples above the Cut-Out Threshold before A/D
Averaging is suspended.
P4.5
Display Update
Rate
0 <
0.25
0.50
0.75
1.00
Configures how often the display is updated in
seconds. If set to 0 the display updates at full
speed.