15
Technical Manual
Zero and Motion Settings Sub-block 2.x
Parameter
Value
Description
P2.0
Pushbutton
Zero Range
2% <
18%
90%
Selects the range (from zero to capacity) within
which the scale can be zeroed. LFT must be 2%.
Example: Scale can be zeroed within ± 2% of
calibrated zero.
P2.1
AZSM Zero
Tracking
OFF
0.5d <
1d
2d
3d
Selects the zero tracking range specified in +/-
displayed divisions. Example: Automatically
zeros the scale within ± 0.5d of calibrated zero.
Must be within the Zero Range.
P2.2
Power Up
ZERO IZSM
0 = Disabled <
1 = Enabled
When enabled, the scale will automatically zero
on power up (Up to 18% of scale capacity).
P2.3
Scale Motion
OFF
1d
2d <
3d
5d
10d
Selects the Scale Motion band in displayed
divisions. Determines the number of divisions at
which the scale is sensitive to motion (Motion
annunciator turns ON).
P2.4
Motion Timer
4 <
Range: 2 - 20
Selects the time (in 0.25 second intervals) the
Motion annunciator will remain ON after the scale
weight stabilizes within tolerance. Example: For
a motion time of 1 second, set this value to 4.
P2.5
Blank Display
on Motion
0 = Disabled <
1 = Enabled
When enabled, the display blanks when motion is
detected.