OPERATING INSTRUCTIONS

Positioning Electrode in a Culture Cup

Proper placement of the electrode in the cell is critical to making accurate 

measurements. The 

STX2-PLUS

 

electrode is designed to facilitate measurements of 

membrane voltage and resistance of cultured epithelia 

in tissue culture wells. The lengths of the electrodes 

are unequal allowing the longer (external) electrode to 

touch the bottom of the dish containing the external 

culture media while preventing the shorter (internal 

electrode) from reaching the bottom of the tissue 

culture cup or insert (see Fig. 1) below). This feature 

ensures proper repeatable positioning between the 

electrode and the cell layer in the cup during the trans 

membrane measurement. The reproducibility of the 

measurements is significantly improved.

Fig. 2— (Right) Electrodes in solution.

Stable Repeatable Readings

This new electrode design sets the correct depth and centering of the electrode 

elements. Once placed and released, movement ceases, the 

EVOM3

 readings become 

more stable, negating any need to make multiple measurements.

Ease of Use

The 

STX2-PLUS 

is designed for use with Corning 347X 24-well plates or Millipore 

hanging inserts. The apical volume is 250 µL, and the basal volume is 750 to 1000 µL.
Insert the outer electrode into the basal access, center the lower adjustment ring onto 

the upper edge of the insert and let the electrode stand on its own. The electrode legs 

no longer need to flex or be inserted at an angle (Fig. 4). The electrode placement is 

stable and repeatable. 
*For 24-well HTS plates, keep the basal electrode centered in the basal access.
*For 12-well inserts, the 

STX2-PLUS

 must be hand held.

Fig. 3—Correct 

Fig. 4—Angle variance

Fig. 5—Depth variance

Содержание STX2-PLUS

Страница 1: ...WORLD PRECISION INSTRUMENTS STX2 PLUS for use with EVOM3 Epithelial Voltohmmeter Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 082620...

Страница 2: ......

Страница 3: ...NG INSTRUCTIONS 3 Positioning Electrode in a Culture Cup 3 Stable Repeatable Readings 3 Ease of Use 3 Adjusting the Length of the Electrodes 4 MAINTENANCE 5 Cleaning Maintaining the STX2 PLUS Electrod...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...rectly in 12 and 24 well tissue culture plates The electrode incorporates a adjustable pair of probes 5 mm tall 4 mm wide and 0 7 mm in thickness Each probe has an outer current and an inner voltage e...

Страница 6: ...any parts are missing at 941 371 1003 or customerservice wpiinc com Returns Do not return any goods to WPI without obtaining prior approval RMA required and instructions from WPI s Returns Department...

Страница 7: ...ility of the measurements is significantly improved Fig 2 Right Electrodes in solution Stable Repeatable Readings This new electrode design sets the correct depth and centering of the electrode elemen...

Страница 8: ...a hanging insert Fig 6 Rotate the nut to adjust the length of the electrodes Fig 7 The electrode length is properly adjusted 2 Align the electrodes with the well inserts and position the STX2 PLUS ove...

Страница 9: ...to your maintenance schedule and the frequency of your cleaning Soak the electrodes overnight when the electrodes have not been on a routine maintenance cleaning schedule Soak 60 minutes if you are o...

Страница 10: ...us electrode must be dipped into chlorine bleach 5 sodium hypochlorite on occasion to maintain the amount of Ag to AgCl balance Otherwise the electrode will become unstable A 10 15 minute immersion of...

Страница 11: ...should show an R2 value 0 99x NOTE These resistance readings can vary due to fluid volume and electrode depth These values are a reference An Excel scatter plot of the molarity vs resistance as a pow...

Страница 12: ...ng Radio frequency interference Turn off or move any cellular phones farther away from the experimental setup NOTE If you have a problem issue with that falls outside the definitions of this troublesh...

Страница 13: ...25 mL of 20 mM KCl 25 mL 50 mL 10 mM KCl APPENDIX B COMPATIBLE INSERTS AND PLATES Corning Millipore Material Membrane Diameter mm Growth Surface Area cm2 Membrane Pore Size m 3470 6 5 0 33 0 4 3472 PI...

Страница 14: ...10 World Precision Instruments Millicell Membrane Pore Size m Qty Pkg MCHT24H48 0 4 48 MCRP24H48 1 0 48 MCSP24H48 3 0 48 MCMP24H48 5 0 48 MCEP24H49 8 0 48...

Страница 15: ...special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or i...

Страница 16: ...Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong New District Shanghai PR China...

Отзывы: