14
World Precision Instruments
APPENDIX: NANOLITER 2000/MICRO4 VOLUME
SETTINGS
When using the
Micro4
to control injections with the
Nanoliter 2000
, take
care when entering the injection volume. The
Nanoliter 2000
injector’s
volume per step is based on the movement of the plunger wire inside a
pulled glass pipette. This plunger moves 0.0005" (12.7µm) for each step of
the motor. The volume of 2.3nL per step is based on the inside diameter of
a 0.5mm pipette and the 12.7µm movement of the plunger wire. As shown
in the table below, use Type K when programming the
Micro4
for use with
the
Nanoliter 2000
.
NOTE: FOR FULL OPERATING INSTRUCTIONS FOR THE MICRO4 AS IT
RELATES TO THE NANOLITER 2000 OPERATION, REFER TO THE UMP3
INSTRUCTION MANUAL.
Table 3: Micro4 Volume Setting Table
Type
Syringe
Volume
Scale
Length
ID (mm)
nL / step REV
K
Max.
Rate
nL/s
Max. Rate
Microstep
Mode nL/s
A
0.5µL
54.1mm
0.1085
0.0294
20
1
B
1.0µL
54.1mm
0.1534
0.0587
40
2
C
5µL
54.1mm
0.343
0.2934
202
14
D
10µL
54.1mm
0.485
0.5868
451
29
E
25µL
60mm
0.73
1.329
1022
66
F
50µL
60mm
1.03
2.646
2035
132
G
100µL
60mm
1.46
5.315
4088
265
H
250µL
60mm
2.3
13.191
9999
659
I
500µL
60mm
3.26
26.501
9999
1325
J
1000µL
4.61
52.995
9999
2649
K
Nanoliter
2000
0.48
plunger
in 0.50
glass
2.3 nL /step
(0.0005” step)
884
115
L
10µL
60mm
0.4607
0.5293
407
29
M N
O P
User
Defined
See page 14
custom
rate
*
**
ILS005
0.4856
0.5880
compensates
for length as
TYPE M
*The rate maximum is calculated internally and is determined by the nL/
step value times 10000 and divided by 13. The microstepping maximum
rate is nL/step*10000/200.
A203XVY
(original
Nanoliter
) uses type M, N, O or P with a value=1.15nL/step.