WPI MPS-4 Скачать руководство пользователя страница 17

MPS-4

World Precision Instruments 

13

MAINTENANCE

Cleaning

Clean the tubing system before and after each experiment. Although the solution path 

of the solenoid valve in 

MPS-4

 does not contact with any metal parts, but the drugs 

remaining in the solenoid valve will affect the accuracy of subsequent experiments. 

So at the end of each experiment, the valve needs to be flushed with distilled water. 

Drain the water out afterwards. The protocols for cleaning are as follows:
1.  Remove the perfusion µ-manifold from the holding rod. 
2.  Carefully remove each of the 8 pieces of tubing from the adaptors on the back 

of the manifold. It is best to push them from their ends since pulling them off 

may damage the perfusion manifold. Pressurizing the tubing may facilitate this 

procedure.

3.  Turn on the control switches and discharge the drug solutions from all 8 tubing 

channels. 

4.  Keeping the switches open, fill each syringe with warm, distilled water to wash the 

tubing and valves. Repeat this step 2 to 3 times.

5.  After the manifold is removed from the tubing (step 2), press the provided 

cleaning adaptor onto the manifold from the tip end. Connect it to a syringe filled 

with filtered water and flush it. 

CAUTION

: Unfiltered water could clog the manifold and permanently damage 

it. The manifold is made of PMMA material. It can only be washed with water. 

Any organic solvent, even alcohol, can permanently damage it. If alcohol must 

be used, only use ethyl alcohol.

TROUBLESHOOTING

If there is no perfusion, check the following to locate the problem:
1.  You can tell if the perfusion controller is running a perfusion sequence by looking 

at the lights above the numbered manual control buttons. When a valve is being 

opened, the corresponding light should turn on. If the perfusion controller won’t 

turn on at all, check the power cord and the fuse in the back panel. If there is 

trouble communicating with the computer in Online mode, make sure the serial 

cable is tight and try restarting both the instrument and the PC software.

2.  The lights above each channel of the valve console turn on when the valve is 

opened. In addition, there a soft click when a valve is turned on or off. If the valve 

console is not responding, tighten the cable that connects it to the perfusion 

controller.

3.  Make sure the stopcock is in the correct position. The middle protrusion on the 

knob should be facing away from the syringe fluid port on the side.

4.  Visually check for air bubbles or obstructions in any segments of tubing. Test 

the µ-manifold by connecting a syringe directly to one of its input passages 

Содержание MPS-4

Страница 1: ...WORLD PRECISION INSTRUMENTS MPS 4 Multichannel Perfusion System Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 011720...

Страница 2: ......

Страница 3: ...Installation 6 Installation 6 Startup 6 Creating a New Perfusion Experiment 6 Set the Experiment Time 7 Preset Each Channel s Perfusion Time 7 Saving Your Experimental Parameters 9 Mode Selection 9 Co...

Страница 4: ...ii World Precision Instruments...

Страница 5: ...nfluence of metal ions in the moving iron of the solenoid valve on the experimental results Unlike other perfusion systems on the market which often compromise performance to fit every possible applic...

Страница 6: ...me is less than 100 nL The injection molded manifold is also designed as an economical disposable item eliminating problems of cross contamination from other experiments Notes and Warnings NOTE Before...

Страница 7: ...0mL Syringes 10 3 way Stopcock 10 Luer fitting with barb for 1 16 ID tubing 5 Color Coded Polyurethane Tubing Ribbon 1 manifold Holding Rod 1 manifold with 100 m ID tip 1 manifold with 250 m ID tip 1...

Страница 8: ...of this manual INSTRUMENT DESCRIPTION Hardware Installation 1 Set the MPS 4 perfusion stand on a stable platform 2 Insert the stainless steel post into the base and tighten the screw 3 Loosen the scre...

Страница 9: ...TING INSTRUCTIONS Fig 3 The front panel of the MPS 4 has the operational controls The left section of the MPS 4 front panel contains the Start Reset and Mode buttons and three functional indicator LED...

Страница 10: ...install an unsigned driver on Windows 8 on page 19 Appendix D Install the MPS 4 software on Windows 10 on page 22 Startup The perfusion software can be found in the Start menu of Windows under C Prog...

Страница 11: ...with the keyboard Move to the next field by filling in both digits of the field or by pressing the spacebar Click Apply when all fields are set to the desired values The perfusion time shown at the ri...

Страница 12: ...el click Apply to verify the changes Otherwise the new values will be lost when a different channel is chosen or the Setup window closes 5 You can also set the perfusion time by holding the left mouse...

Страница 13: ...der to change the operation mode or choose the serial port choose Mode Setup from the Operation menu press the F3 function key or click the button in the Tool Box The Mode Setup window pops up Click o...

Страница 14: ...on The button runs the perfusion process In the Online mode it acts as the button and runs the experiment controlled by the perfusion software In Offline mode the Start button runs the preset perfusio...

Страница 15: ...de Press any of the Channel keys if the experiment has to be manually halted This will exit out of TTL Control mode and prevent it from going back into that mode until Reset is pressed or the instrume...

Страница 16: ...t the tip of the micro perfusion head is close to the bottom of the dish The access angle is about 35 45 5 Pull a 1 m micropipette Fill the pipette with 150mM NaCl solution to make it a microelectrode...

Страница 17: ...lush it CAUTION Unfiltered water could clog the manifold and permanently damage it The manifold is made of PMMA material It can only be washed with water Any organic solvent even alcohol can permanent...

Страница 18: ...210x170 mm APPENDIX A DETERMINING FLOW RATE Theoretical Calculation The relationship of flow rate to the height of the liquid column and inner diameter of the capillary tubing can be accurately predic...

Страница 19: ...driver Please follow those steps below 1 Since you need to restart the computer multiple times during the operation you need to save your unfinished work and exit all applications before the disable d...

Страница 20: ...uments 3 Click the option Update security and a new window appears Fig 14 Select the Recovery tab 4 After selecting the Recovery tab the Advanced startup area appears on the right side of the window F...

Страница 21: ...restart 6 After restart the Option screen displays Fig 16 Click Troubleshoot 7 After clicking the option Troubleshoot the Troubleshoot screen appears Fig 17 Click Advanced Options 8 Click the option...

Страница 22: ...again The following menu appears on the screen after the restart Fig 20 Press 7 on your keyboard 11 You can press 7 on your keyboard to choose the Disable driver signing enforcement The computer will...

Страница 23: ...low the steps shown below to start Windows 8 in Disable Driver Signature Enforcement mode 1 Move the cursor over the top or bottom right corner or the screen and it will show extra options on the righ...

Страница 24: ...n the left hand side and then click Restart now button on the right Fig 23 Select the General tab and click Restart now 4 After a restart click Troubleshoot Fig 24 Select the Troubleshoot option 5 Cli...

Страница 25: ...Fig 27 Press Restart 8 Choose the Disable Driver Signature Enforcement and hit the Enter key to start Windows Fig 28 Select the Disable Driver Signature Enforcement option and press Enter 9 You shoul...

Страница 26: ...e click on it to begin the installation 3 Using the provided USB cable connect the USB port of host computer to the USB port square of the MPS 4 controller and then turn on the power of the MPS 4 cont...

Страница 27: ...on your computer depending on what is installed on your computer When the device manager is opened the Other devices Unknown device with yellow background exclamation mark will appear in the director...

Страница 28: ...nts 7 Click the Update Driver button The following window appears Fig 33 Select the option to Browse my computer 8 Select and click Browse my computer for driver software A similar window appears Fig...

Страница 29: ...from running MPS 4 19 07 exe during the installation For example on the this computer the MPS 4 program was specified to be installed in the following directory location C Program Files x86 MPS 4 Dri...

Страница 30: ...ver software was installed from running MPS 4 19 07 exe in step 2 Click the Next in the lower right corner and the following window appears Fig 38 Click the Install option 13 Select and click the Inst...

Страница 31: ...ontrollers section and confirm a successful installation of the driver software NOTE After the installation of the MPS 4 driver software a system reboot automatically resumes the state of the driver s...

Страница 32: ...28 World Precision Instruments...

Страница 33: ...MPS 4 World Precision Instruments 29...

Страница 34: ...30 World Precision Instruments...

Страница 35: ...ecial or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or ind...

Страница 36: ...n Germany Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong Free Trade Zone Shang...

Отзывы: