8.6 Service By the Approved Technician
Service is essential to effective sterilization and continued running.
We suggest general servicing by an approved technician per 2-year, or 2500 cycles.
Every 3 months replacing the bacteriological filter, every year replacing the door seal.
Check-list based on general servicing:
1. Check the solenoid valves
2. Check the water pump
3. Check the vacuum pump
4. Check the distilled water drain valve and the used water drain valve
5. Check the safety valve
6. Check the door-locking system
7. Check the sensor of pressure and temperature
8. Check the water level sensor
9. Check the electrical connections
10. Check the hydraulic connections
11. Check the safety thermostat
12. Clean the sterilization chamber
13. Clean the trays and the tray holder
14. Clean the reservoirs
15. Replace the water filter
16. Replace the air filter
17. Replace the door gasket
Statement: Under recommended operating conditions and maintenance, calibration
is not required.