Steam Sterilizer Operation Manual
TANZO Classic Model Page of 46
36
Clean After Liquid Loads
Biological media tends to boil at a higher rate than other liquids during venting. This
causes media to be spattered inside the chamber. Therefore, the chamber must be
cleaned daily when you are sterilizing media. Cleaning as follows:
❖
Allow unit to be cooled down.
❖
Wipe out chamber and door with a clean, damp cloth.
WARNING
Failure to keep the interior of the stainless steel chamber free of
mineral deposits and debris can cause premature failure of the
sterilizer.
8.3 Weekly Maintenance (More Often If Necessary)
Clean Chamber, Trays and Rack
At least once a week, the trays and tray rack should be removed from the sterilizer
chamber. The trays, tray rack and chamber should be thoroughly cleaned to remove any
deposits from the surfaces.
Clean the trays, rack and chamber (especially the bottom of the chamber) with
appropriate antibiological cleaners. Wipe all residues from the surfaces with a dampened,
lint-free cloth.
WARNING To prevent from collection of mineral deposits and corrosion of
chamber components, use distilled or deionized water only as
specified. Clean chamber after each use if sterilizing saline solutions.
Clean Water Filter
pic 8-1
Water draining filter might has been jammed by some dust because of use for a long-term,
so effect of vacuum and drying would be influenced. Some tiny impurity might be
deposited on the filter after a long-term use, blocking the filter, so as to influence the effect
of the vacuuming and water discharging. The kinds of impurity come from smeary dust on