Steam Sterilizer Operation Manual
TANZO Classic Model Page of 46
37
the instruments being sterilized or some calcification in the water.
Keep cleaning of the inside chamber in order to make life-time of filter much longer;
please take the following advice for consideration:
❖
Use eligible distilled water;
❖
The instruments should be cleaned before placing in; it is good to use specified
packing
for the instruments with oil or other impurity, don’t forget to seal up.
❖
Rotate the water filter which composed by filter net tube(A) and filter holder(B) inside
the chamber. Cleaning the part A and B, ensure that there is no any dirty thing on
it.( We suggest you to clean by ultrasonic cleaning machine). Then set it back and
rotate to the hold bottom of the chamber.
(
pic 8-2
)
8.4 Monthly Maintenance
Clean Water Tank
There are some impurities and some toxins had been left behind in the reservoir because
distilled water stored for a long time, so you need to drain and clean regularly. According
to the picture which shows below, loose the screw by screw-driver, and open the cover to
clean inside. As pic 8-3 shows
(
pic 8-3)
ATTENTION BE SURE THAT USE DISTILLED WATER PROPERLY IN ORDER TO
EXTEND THE STERILIZER LIFE-TIME.
DO NOT RAVE ABOUT THE STERILIZER WHEN THE TANK HAS
BEEN FILLED.
A
B