Steam Sterilizer Operation Manual
TANZO Classic Model Page of 46
29
If the water shortage gives an alarm during the sterilizing, you should not worry about it,
the water will not influence this process. And the adequate water should be left in the tank
for the operating cycle, but you need to add water in time to ensure the next operation.
6.3 Alarm if Used Water Reservoir is Full
The
“
OUT
” indicator light illuminates
during sterilizing, which means that the used water
reservoir needs to be drained out.
Connect the water tube to the machine left lower corner inlet connector as followed picture
arrow marked.
Generally, the maximum temperature of drained water should be under 70
℃
. If it is higher,
you need to check whether the fan operates normally, or contact the local distributor
immediately, we will offer our best service to you in no time.
6.4 Select Program
Select the required sterilization program and temperature which you need. When you
choose, the corresponding indicator light will illuminate. After the sterilization program
being selected, the temperature will be automatically set as default model. It is
recommended that the sterilization program should not be changed except for special
process.
6.5 Load Articles
Articles should be placed on the trays with some gap among them, so that the steam can
CAUTION USE DISTILLED WATER ONLY TO EXTEND THE LIFE TIME OF THE
MACHINE.
DO NOT TILT THE STERILIZER WHEN THE RESERVOIR IS FULL OF WATER.
Clean-Water out
Used-Water out