1.7 Device safety classification

1.7.1 Type of operation mode: Continuous operating device
1.7.2 Type of protection against electric shock: Class II equipment with 
internal power supply
1.7.3 Degree of protection against electric shock: BF type applied part
1.7.4 Degree of protection against harmful ingress of water: Ordinary 
equipment (IPX0)
1.7.5 Degree of safety application in the presence of a flammable 
anesthetic mixture with air, oxygen, or nitrous oxide: Equipment cannot 
be used in the presence of a flammable anesthetic mixture with air, 
oxygen, or nitrous oxide.
1.7.6  Applied part: contra-angle.
1.7.7  The contact duration of applied part: 1 to 10 minutes.

1.7.8  The temperature of the surface of applied part may reach 46.6℃.

1.8 Primary technical specifications

1.8.1 Battery  
         Lithium battery in handpiece: 3.6V /750mAh
1.8.2 Power adapter
         Input: ~100V-240V 50Hz/60Hz 0.4A Max
         Output: DC5V/1A
1.8.3 Torque: 0.6Ncm-5.0Ncm(6mNm ~ 50mNm)
1.8.4 Rotate speed: 100rpm~1000rpm

1.9 Environment parameters

1.9.1 Environment temperature: +5℃ ~ +40℃

1.9.2 Relative humidity: 30% ~ 75%
1.9.3 Atmospheric pressure: 70kPa ~ 106kPa

2 Installation

2.1 Basic accessories of product

            

                 

Содержание E-COM

Страница 1: ...E COM Endo Motor INSTRUCTION MANUAL...

Страница 2: ...ion of product 7 4 Operation instruction 8 5 Oiling of contra angle 14 6 Cleaning Disinfection and Sterilization 14 7 Troubleshooting 23 8 Storage maintenance and transportation 24 9 Environmental pro...

Страница 3: ...uring root canal preparation procedure it is used to mold and clean the root canal Features a Wireless handpiece enables more convenient operation b Adopt real time feedback technology and dynamic tor...

Страница 4: ...arned not to use household appliances such as electric razors hair dryers etc are not recommended to use this device 1 6 Warnings 1 6 1 Please carefully read this Instruction Manual before first opera...

Страница 5: ...n damage to lithium battery and control circuit 1 6 9 The handpiece cannot be autoclaved Use disinfectant of neutral pH value or ethyl alcohol to wipe its surface 1 6 10 Before the contra angle stoppi...

Страница 6: ...annot be used in the presence of a flammable anesthetic mixture with air oxygen or nitrous oxide 1 7 6 Applied part contra angle 1 7 7 The contact duration of applied part 1 to 10 minutes 1 7 8 The te...

Страница 7: ...ect contra angle with disinfectant of neutral PH value After disinfection lubricate it with specific cleaning oil Finally sterilize it under high temperature and high pressure 134 2 0bar 2 3bar 0 20MP...

Страница 8: ...stallation of file Before starting the device plug the file into the hole of contra angle head While plugging slightly screw the file with one hand and press the push cover of contra angle with anothe...

Страница 9: ...s cleaned and disinfected put on a disposable isolation sleeve Take the isolation sleeve out of the isolation sleeve box then insert the isolation sleeve into the motor handpiece from the thin end of...

Страница 10: ...consumption d Set speed e Set torque f File system g Operation mode h Torque setting i Real time torque j Set apical stop k File position 4 Operation instruction 4 1 Starting and Stopping a Under the...

Страница 11: ...to switchover those memorizing programs And the serial number will change correspondingly 4 3 Speed setting torque setting and operation mode setting Under standby state short press key to enter spee...

Страница 12: ...or key to switch operation mode Continuous Rotation Mode Reverse Rotation Mode and Reciprocating Motion Mode are for options There will be a tick indication while setting to Reverse Rotation Mode Long...

Страница 13: ...W3 ONE W3 Single Press key to get into the interface of corresponding file system Press key or key to select needed file model the corresponding rotation direction speed and torque value Press the ke...

Страница 14: ...r charging please unplug the power adapter Note Please use the original power adapter for charging 4 6 Contra angle calibration setting After replacement of contra angle the contra angle shall be cali...

Страница 15: ...wise rotation Load value is lower than preset torque value b Counterclockwise rotation Load value is higher than preset torque value c Clockwise rotation Load value is lower than preset torque value a...

Страница 16: ...contra angle for 2 3s till the oil flow out of contra angle head part 3 Vertically place the end part of contra angle or tilt the contra angle to let go the redundant oil under gravity Warnings Handp...

Страница 17: ...cable after sterility 6 2 7 The water quality has to be convenient to the local regulations especially for the last rinsing step or with a washer disinfector 6 2 8 To sterilize the endodontic files re...

Страница 18: ...fection must be performed within 10min before use b The disinfectant used must be used immediately no foaming is allowed c In addition to 75 alcohol you can use non residue disinfectants such as Oxyte...

Страница 19: ...acturer takes no responsibility The products may not be exposed to temperature above 138 Processing limit The products have been designed for a large number of sterilization cycles The materials used...

Страница 20: ...ater used here must be pure water distilled water or deionized water 6 4 2 Preparation before cleaning Steps Tools tray soft brush clean and dry soft cloth 1 Remove the shanks files 2 Remove the isola...

Страница 21: ...e please refer to the automated disinfection section in the next section Disinfection Notes a The cleaning agent does not have to be pure water It can be distilled water deionized water or multi enzym...

Страница 22: ...e should be suitable for the product to be treated The flushing period should be sufficient 5 10 minutes Pre wash for 3 minutes wash for another 5 minutes and rinse it for twice with each rinse lastin...

Страница 23: ...he drying temperature should not exceed 138 C c The equipment used should be inspected and maintained regularly 6 4 6 Inspection and maintenance 6 4 6 1 Inspection In this chapter we only check the ap...

Страница 24: ...zation procedures are prohibited The steam sterilizer complies with EN13060 or is certified according to EN 285 to comply with EN ISO 17665 The highest sterilization temperature is 138 C The steriliza...

Страница 25: ...an atmospheric pressure of 70KPa to 106KPa and a temperature of 20 C to 55 C 2 After sterilization the product should be packaged in a medical sterilization bag or a clean sealing container and store...

Страница 26: ...distributor or manufacturer The continuously rotating file is stuck at the root canal Incorrect specification setting Too high load torque of file Choose Reverse Rotation Mode start the handpiece and...

Страница 27: ...8 3 3 Avoid solarization and getting wet in rain and snow during transportation 9 Environmental protection Please dispose according to the local laws 10 After service From the date this equipment has...

Страница 28: ...erved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner s...