and disinfected with automatic equipment. Manual cleaning and
disinfection is required.
6.3.1.1 Manual cleaning steps:
1. Take out the handpiece, charger, and base on the workbench.
2. Wet the soft cloth completely with distilled water or deionized water,
and then wipe all the surfaces of the components such as the handpiece,
charger, base, etc. until the surface of the component is not stained.
3. Wipe the surface of the component with a dry soft nap-free cloth.
4. Repeat the above steps at least 3 times.
Note:
a)Use distilled water or deionized water for cleaning at room
temperature.
6.3.1.1 Manual disinfection steps:
1. Soak the dry soft cloth with 75% alcohol.
2. Wipe all surfaces of headpiece, charger, base and other components
with a wet soft cloth for at least 3 minutes.
3. Wipe the surface of the component with a dry soft nap-free cloth.
Note:
a) The cleaning and disinfection must be performed within 10min
before use.
b) The disinfectant used must be used immediately, no foaming is
allowed.
c) In addition to 75% alcohol, you can use non-residue disinfectants
such as Oxytech from Germany, but you must respect the concentration,
temperature and time specified by the disinfectant manufacturer.
d) After cleaning and disinfecting the handpiece, you must install a
disposable isolation sleeve before use and repeat steps 1, 2 and 3 to
clean the disposable isolation sleeve(For detailed installation steps, see
section 2.7).
6.3.2 Post-Op processing
After each use, clean and disinfect the handpiece, charger, and base
within 30 minutes. The specific steps are as follows:
Tools: Nap-free soft cloth, tray
1. Remove the contra-angle from the handpiece, place it in a clean tray,
and then remove the disposable isolation sleeve from the handpiece.
2. Soak the nap-free soft cloth with distilled water or deionized water,
and then wipe all the surfaces of the components such as the handpiece,
charger, base, etc. until the surface of the component is not stained.