Cleaning and disinfecting steps by using Washer-disinfector
1. Carefully place the product into the disinfection basket. Fixation of
product is neededonly when the product is removable in the device. The
products are not allowed to contact each other.
2. Use a suitable rinsing adaptor, and connect the internal water lines to
the rinsing connection of the washer-disinfector.
3. Start the program.
4. After the program is finished, remove theproductfrom the washer-
disinfector, inspect (refer to section "Inspection and Maintenance") and
packaging (refer to chapter "Packaging"). Dry the productrepeatedly if
necessary (refer to section "Drying").
Notes:
a) Before use, you must carefully read the operating instructions
provided by the equipment manufacturer to familiarize yourself with
the disinfection process and precautions.
b) With this equipment, cleaning, disinfection and drying will be carried
out together.
c) Cleaning: (c1) The cleaning procedure should be suitable for the
product to be treated. The flushing period should be sufficient (5-10
minutes). Pre-wash for 3 minutes, wash for another 5 minutes, and rinse
it for twice with each rinse lasting for 1 minute. (c2) In the washing
stage, the water temperature should not exceed 45 °C, otherwise the
protein will solidify and it is difficult to remove. (c3) The solution used
can be pure water, distilled water, deionized water or multi-enzyme
solution, etc., and only freshly prepared solutions can be used. (c4)
During the use of cleaner, the concentration and time provided by
manufacturer shall be obeyed. The used cleaner is neodisher MediZym
(Dr. Weigert).
d) Disinfection: (d1) Direct use after disinfection: temperature ≥ 90 ° C,
time ≥ 5 min or A0 ≥ 3000;
Sterilize it after disinfection and use: temperature ≥ 90 ° C, time ≥ 1
min or A0 ≥ 600
(d2) For the disinfection here,the temperature is 93 ° C, the time is 2.5
min, and A0>3000
e) Only distilled or deionized water with a small amount of
microorganisms (<10 cfu/ml) can be used for all rinsing steps.
(For example, pure water that is in accordance with the European
Pharmacopoeia or the United States Pharmacopoeia).