18
6.4.1.2Post-operative treatment
The post-operative treatment must be carried out immediately, no later
than 30 minutes after the completion of the operation. The steps are as
follows:
1. Remove the products from the base, and rinse away the dirt on the
surface of handpiece with pure water (or distilled water/deionized
water);
2. Dry the products with a clean, soft cloth and place it in a clean tray.
Notes:
a) The water used here must be pure water, distilled water or deionized
water.
6.4.2 Preparation before cleaning
Steps:
Tools: tray, soft brush, clean and dry soft cloth.
1. Remove the shanks/files.
2. Remove the isolation sleeve and Contra-angle from the handpiece,
and then put them into a clean tray;
3. Use a clean soft brush to carefully brush connecting protective
silicon cover, head and back cover of the contra-angle until the dirt on
surface is not visible. Then use soft cloth to dry the products and put
them into a clean tray. The cleaning agent can be pure water, distilled
water or deionized water.
Disassembling steps