13
Veterinary Lactate Meter
Coding the Meter
IMPORTANT:
After setting the date and time for the first time, the current code will appear
on the screen every time the meter is turned on. Verify that the code on the
screen matches the code number on the package of test strips before each use
of the meter. The code needs to be set only once for each package of test strips.
The meter will memorise the code until it is changed. If the code number on the
meter screen doesn’t match the code number on the test strip package, it will
create an inaccurate blood lactate test result.
To code the meter with a code card, follow these steps.
Step 1: Locate the code card in the test strips package.