Veterinary Lactate Meter
14
NOTE
:
There is a code card packaged in each box of test strips. The code card is designed
specifically for use with the test strips in that particular package. The code card
should have the same code number indicated on the package of test strips.
Step 2: Verify that the code number on the code card matches the code number
on the test strip package.
Step 3: Insert the code card with the code number facing up firmly and
completely into the code card port on the back of the meter.
Step 4: Press any button to power on the meter.
Step 5: A 4-digit code number will display on the screen. Verify the code number
on the screen matches with the code card number and the number on the
test strip package.