![Woodley Clinispin MPC Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/woodley/clinispin-mpc/clinispin-mpc_operators-manual_995060008.webp)
page 7
SETUP LOCATION
1) The MPC is supplied in a box. Open the box, then remove the packaging and
gently take the centrifuge out of the small box. Before 1st time usage, open the
centrifuge & ensure to remove all packaging from the rotor chamber & ensure
rotor is firmly tightened. The user manual and accessories should be kept with
the centrifuge.
2) Choose a setup location which meets the following criteria:
a) A bench top clearance height of 40 cms is required in order to open the lid.
b) The clearance envelope is the space around the centrifuge which is
required for safety. Choose a setup location which will allow for a clearance
envelope of at least 60 x 60 cms, (with centrifuge at the centre). No person
or hazardous material shall be permitted in the clearance envelope during
operation. The operator time within the envelope shall be limited to the time
necessary for loading, uploading and centrifuge operation only.
c) Proper ventilation is necessary to prevent the overheating of samples as
well as premature failure of the centrifuge. Choose an area which will allow
unencumbered air flow.
d) The centrifuge is designed to rest on its four rubber feet. No adjustment is
necessary for leveling the centrifuge, however, the surface should be flat
and leveled.
e) Be sure the outlet is always within reach as the main lead is the means of
emergency disconnection.
INITIAL SETUP PROCEDURE
1. Connect one side of power adaptor to rear
side of centrifuge and other to approved
electrical outlet.
2. Rotate clockwise the adaptor nut to tighten the
adaptor with the centrifuge. Ensure the power
switch is OFF while connecting the power adaptor.
Содержание Clinispin MPC