CARE AND MAINTENANCE
With proper care and maintenance your centrifuge will provide years of laboratory
service.
For proper care, the following steps should be taken:
1. Provide Adequate Ventilation:
For cooling purposes, the centrifuge draws in
ambient air through the air intake cover on the top of the lid and exhausts this air in
the rear of the base. The centrifuge should be placed on a hard smooth surface for
good air circulation.
2. Motor and Electrical Maintenance:
This centrifuge uses a brushless-DC motor.
There are no brushes to replace and it should not need routine servicing for the life
of the centrifuge. The electrical components are selected for high reliability and
should not need service.
3. Keep the Centrifuge Clean:
The cabinet, rotor top and accessories shall be
thoroughly cleaned using either soap and water, or a mild bleach solution. If it is
necessary to remove the rotor for additional cleaning it is required that a qualified
technician remove the outside housing and rotor assembly. Contact Woodley
Equipment Company for additional information. Apply cleaning solutions with a
towel or cloth. Do not submerge the centrifuge in water or other cleaning solutions as
this will cause damage and void your warranty!