TROUBLESHOOTING:
Solution: •
Check mains power supply is available.
•
Check that power adaptor is working properly and it is properly
connected at both ends.
Solution: •
Rotor might be spinning. Check in the view window for the rotor
to stop.
•
There might be a power failure. Check in the view window for the
rotor coming to a stop and use Emergency lid release to open.
Solution: •
Rotor not loaded symmetrically. Load the rotor symmetrically
and restart the operation.
• A broken tube or damage to rotor. Clean the rotor chamber if
there is any broken tube or replace rotor if damaged.
• Motor might be causing excessive noise. Contact service
representative.
Solution: •
Rotor not loaded symmetrically / imbalance detected. Restart
centrifuge and load rotor symmetrically.
Solution: •
Rotor jammed - remove any obstruction to allow free rotor movement.
• Motor jammed or damaged. Contact service representative.
Solution:
•
If the temperature of motor is High and reaches beyond 65
degree C. Allow the motor to cool down.
Once motor is cool the error will disappear.
Solution:
•
Electronics error. Switch off the centrifuge and then switch it ON
again. If error still shows, contact service representative.
1. Problem:
No Display
2. Problem:
Centrifuge lid cannot be opened
3. Problem:
Centrifuge shakes during acceleration & excessive noise
4. Problem:
Err 55
5. Problem:
Err 52
For additional assistance, services and technical support
contact Woodley Equipment Company LTD.
7. Problem:
System stops working / freezes