SAFETY PRECAUTIONS

• 

Never use the centrifuge in any manner not specified in this manual.

 Equipment used in any manner not specified in this manual or by the 

  manufacturer, the protection provided by the equipment may be impaired. 

 Never move the centrifuge while the rotor is spinning.

 The rotor and the rotor lid must always be securely fastened. If the centrifuge 

  makes unusual noise during operation, the rotor or rotor lid fitment needs to be 

  checked. Switch off the device immediately by pressing STOP, check rotor lid & 

  fasten securely.

 The rotors must be loaded symmetrically. Each tube should be counter balanced 

  by another tube of same weight and volume.

 Do not use centrifuge or rotor that have not been correctly installed or damaged.

 Repairs must only be performed by authorised service technician.

 Only use recommended original rotors and spare parts.

 Centrifuge may be used for the specified applications only. It must not be 

operated in a hazardous or flammable environment and must not be used to 

centrifuge explosive or highly reactive substances.  Also do not place the 

potential hazardous material within the clearance area/envelope.

 If liquids are spilled on the rotor or rotor chamber, the centrifuge must be cleaned 

  carefully and properly before being used again.

 Prior to centrifugation, the tubes should be visually inspected for material 

  damage. Damaged tubes may not be centrifuged. This is because broken tubes 

  can, in addition to sample loss, create imbalance which can result in further 

  damage to the centrifuge and accessories.

 Rotors and rotor lids are high-graded components which are subject to extreme 

  mechanical strain. Even slight scratches and tears can lead to serious internal 

  material damage. Ensure to check rotor for any signs of damage before use. 

• 

Rotor & rotor lid showing visible signs of corrosion or mechanical damage should 

  not be used. 

page 6

Содержание Clinispin MPC

Страница 1: ...Operator s Manual CLINISPIN MPC Multi Purpose Centrifuge Rev 1 0...

Страница 2: ......

Страница 3: ...up Location Initial Setup Procedure Operation Rotor Configuration Filling of Capillary Tubes Rotor Installation Control Panel Functions Advanced User Settings Care and Maintenance Troubleshooting page...

Страница 4: ...MODEL DESCRIPTION The Clinispin Multi Purpose Centrifuge MPC is a high speed centrifuge equipped with a maintenance free drive digital display and simple interface for silent efficient operation It i...

Страница 5: ...bient Temperature during operation 4 C to 40 C Maximum Relative Air Humidity 80 Noise Level 60 dB Electrical Requirements Input Voltage 110 240 VDC 50 60 Hz Output Voltage 24VDC 6 25 A Power Consumpti...

Страница 6: ...d Button RPM indicator LED Increment button Serum Plasma indicator LED Serum Plasma button preset HCT QBC button preset Urine button preset HCT QBC indicator LED Speed display Time Display RCF indicat...

Страница 7: ...rs and spare parts Centrifuge may be used for the specified applications only It must not be operated in a hazardous or flammable environment and must not be used to centrifuge explosive or highly rea...

Страница 8: ...s material shall be permitted in the clearance envelope during operation The operator time within the envelope shall be limited to the time necessary for loading uploading and centrifuge operation onl...

Страница 9: ...trifuge A smooth whirring sound should be heard If there are any loud or unusual sounds stop the centrifuge by pressing the OPEN STOP button immediately and refer to the section on troubleshooting pag...

Страница 10: ...d IF A PROBLEM IS FOUND DURING A SPIN THAT REQUIRES THE CENTRIFUGE TO BE SHUT DOWN PRESS THE OPEN STOP BUTTON IMMEDIATELY 8 Once the set speed is reached the timer will begin the remaining time will b...

Страница 11: ...proper tube loading 1 Opposing tube carriers must be identical 2 Opposing tube carriers must be empty or loaded with an equal number of equally weighted samples 3 If an odd number of samples is to be...

Страница 12: ...nd of the capillary Insert the capillary tube vertically into the seal up to the point where the end of the capillary touches the bottom of the seal plate Slightly tilt the capillary tube to the side...

Страница 13: ...d load vertically on the motor shaft 2 Place the rotor nut in the center hole of the rotor onto the motor shaft 3 Put T Allen Key in rotor nut turn clockwise to tighten and anti clockwise to loosen ro...

Страница 14: ...HCT QBC For HCT rotor This HCT QBC button has the pre defined speed time value of 12000 RPM 5 minutes respectively Single press the HCT QBC button to start operation at 12000 RPM for 5 minutes of tim...

Страница 15: ...will run for a maximum of 99 minutes and will then shut off automatically The ongoing centrifugation may also be terminated by pressing the STOP OPEN button The remaining time after the termination o...

Страница 16: ...the centrifuge lid opens release the emergency release cap back in the hole Emergency lid release pin Pull Down using any tool Lid unlock Imbalance Detection This centrifuge is equipped with imbalance...

Страница 17: ...here are no brushes to replace and it should not need routine servicing for the life of the centrifuge The electrical components are selected for high reliability and should not need service 3 Keep th...

Страница 18: ...epresentative Solution Rotor not loaded symmetrically imbalance detected Restart centrifuge and load rotor symmetrically Solution Rotor jammed remove any obstruction to allow free rotor movement Motor...

Страница 19: ......

Страница 20: ...acement parts or accessories Old Station Park Buildings St John s Street Horwich Bolton BL6 7NY United Kingdom Tel 44 0 1204 669033 Fax 44 0 1204 669034 www woodleyequipment com Email sales woodleyequ...