Emergency Lid Release
Disconnect the centrifuge from the main supply. Wait until the rotor has come to a
standstill (this may take a longer time). Once the rotor has stopped, use any tool
to pull down the emergency lid release cap. Insert the tool in the hole of emergency
lid release cap and push down. This will open the centrifuge lid. After the centrifuge
lid opens, release the emergency release cap back in the hole.
Emergency
lid release pin
Pull Down
using any tool
Lid unlock
Imbalance Detection
This centrifuge is equipped with imbalance detection for user safety. In case of major
imbalance, the centrifuge will terminate the current run and will begin to brake to a
stop. The words “ERR 55” will flash on the speed display. Once the rotor has stopped,
restart the centrifuge using
‘POWER’ button to cancel the error reporting, balance
the
the load and begin a new run.
ADVANCED USER SETTINGS :