Naming the Cycle

 

 

Navigate to the cycle name with the left and right arrow. Change 
the blinking character of the name with the up and down buttons, 
then move to the next character with the right arrow. Repeat.

 

 

 

 

Save and Exit 

Programming Mode

 

 

Press the GEAR button. The cycle displayed on the screen will be 
automatically saved.

 

 
 

S

ETTINGS 

 

 

(1) Chemistry 

(2) Coag (PPP) 

(3) Urine 

 

RPM 

3,600 

3,300 

1,900 

HORIZON 6 

Flex 

Time 

10 

15 

G-force 

1,800 xg 

1,500 xg 

500 xg 

 

Preset Cycle 

10 min @ 3,900 RPM 

15 min @ 3,800 RPM 

5 min @ 2,200 RPM 

HORIZON 6 

Flex FA 

75 & 100 mm tubes 

1,600 xg 

1,500 xg 

500 xg 

125 mm tubes 

1,850 xg 

N/A 

600 xg 

 

 

O

PERATION 

o

 

Place the tubes into the tube holders. Be sure to follow the rules for balanced loads as listed in the next section.

 

o

 

Close the lid and turn the lid knob clockwise to its complete stop position.

 

o

 

The digital screen shows the currently selected cycle. To select another cycle, press the UP or DOWN button in 
succession until the desired cycle is selected.

 

 

o

 

Pushing the START button on the control panel will start the spin cycle.

 

o

 

When the cycle is completed, the rotor will slow to a complete stop and the lid light will flash.

 

o

 

The unlocking mechanism will engage for 60 seconds allowing entry into the rotor chamber. To unlock after more 
than 60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds.

 

 

o

 

Turn the lid knob counterclockwise and open the lid. The lid light will turn off.

 

Содержание Clinispin Horizon 6 Flex

Страница 1: ...Clinispin HORIZON 6 Flex 6 Flex FA Operator s Manual...

Страница 2: ......

Страница 3: ...1 Initial Setup 2 Quick Start 3 Single Cycle Lock 4 Preset Lock 4 Setting or Modifying a Cycle 4 Settings 5 Operation 5 Balancing Loads 6 Care and Preventative Maintenance 7 Cleaning and Disinfection...

Страница 4: ...verheating of samples by using ambient air to keep specimens at room temperature o The tube holders are fibre reinforced for high strength durability and years of trouble freeuse o A clear lid permits...

Страница 5: ...and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be...

Страница 6: ...cycle displayed on the screen The lid must be closed Unlock Allows access into the rotor chamber by engaging the unlocking mechanism Entry is only possible when the rotor is stopped Stop Pressing the...

Страница 7: ...es can again be edited or programmed SETTING OR MODIFYING A CYCLE Up to 10 cycles can be named programmed for time speed and braking and saved in memory Cycle Selection Press the up and down buttons n...

Страница 8: ...ubes 1 850 xg N A 600 xg OPERATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise...

Страница 9: ...e life of the centrifuge and produce better results Use the following rules when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired...

Страница 10: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Страница 11: ...er with a tool to remove them o Inspect the rotor tube holders or buckets for damage o If the tube holders or buckets have any damage evenslight safely dispose of them and replace them o If the rotor...

Страница 12: ...CK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use apen to manually disengage the locking mechanism...

Страница 13: ...2 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 220 Watts 220 Watts Centrifuge Motor H P Brushless H P Brushless Max G force 2 000 xg 1 600 xg 1 850 xg Max Speed...

Страница 14: ...e of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipm...

Отзывы: