NOTE: The latch must be turned completely clockwise to its stop position for the centrifuge to operate.

 

 
 

 

The centrifuge does not run

 

 

o

 

Verify that the centrifuge is powered. The screen should be lit.

 

o

 

If “Lid not closed” message is displayed, make sure the lid latch 

is turned completely clockwise to its stop position.

 

o

 

If the centrifuge still does not run, contact Technical Support.

 

The rotor does not spin freely 

 

o

 

Make sure nothing has fallen into the rotor chamber, 
following the procedure above.

 

o

 

If nothing obstructs the rotor, the rotor may be damaged. 
Contact Technical Support for further assistance.

 

The centrifuge makes a rattling noise 
when running 

o

 

Stop the centrifuge. Open the lid.

 

o

 

Wearing PPE, remove tubes and tube holders/buckets and 
look for fallen objects or debris. Carefully reach inside the 
rotor chamber with a tool to remove them.

 

o

 

Inspect the rotor, tube holders or buckets for damage.

 

o

 

If the tube holders or buckets have any damage, even slight, 
safely dispose of them and replace them.

 

o

 

If the rotor appears damaged, contact Technical Support 
for further assistance.

 

Excessive noise or vibration when the 
centrifuge is running

 

 

o

 

Verify that all four centrifuge feet are properly seated on a flat 
surface.

 

o

 

Ensure that the load is balanced according to instructions in 

the “Balancing Loads” section of this

 manual.

 

o

 

Make sure that nothing has fallen into the rotor chamber.

 

The  centrifuge  stops  and  beeps 
continuously 

 

The load is not balanced. Press the UNLOCK button, open the lid, 
and balance the load as recommended on page 6 of this manual. 

The centrifuge is stuck on one of the 
settings

 

 

Cycle selection is locked. Press the UNLOCK button for 5 seconds.

 

The cycle time and speed are not set to 
the desired value

 

 

Check the setting by following the instructions in the section on 
Setting or Modifying a Cycle. If the preset is not the desired length, 
follow the procedure on the same page to change the preset time.

 

Содержание Clinispin Horizon 6 Flex

Страница 1: ...Clinispin HORIZON 6 Flex 6 Flex FA Operator s Manual...

Страница 2: ......

Страница 3: ...1 Initial Setup 2 Quick Start 3 Single Cycle Lock 4 Preset Lock 4 Setting or Modifying a Cycle 4 Settings 5 Operation 5 Balancing Loads 6 Care and Preventative Maintenance 7 Cleaning and Disinfection...

Страница 4: ...verheating of samples by using ambient air to keep specimens at room temperature o The tube holders are fibre reinforced for high strength durability and years of trouble freeuse o A clear lid permits...

Страница 5: ...and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be...

Страница 6: ...cycle displayed on the screen The lid must be closed Unlock Allows access into the rotor chamber by engaging the unlocking mechanism Entry is only possible when the rotor is stopped Stop Pressing the...

Страница 7: ...es can again be edited or programmed SETTING OR MODIFYING A CYCLE Up to 10 cycles can be named programmed for time speed and braking and saved in memory Cycle Selection Press the up and down buttons n...

Страница 8: ...ubes 1 850 xg N A 600 xg OPERATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise...

Страница 9: ...e life of the centrifuge and produce better results Use the following rules when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired...

Страница 10: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Страница 11: ...er with a tool to remove them o Inspect the rotor tube holders or buckets for damage o If the tube holders or buckets have any damage evenslight safely dispose of them and replace them o If the rotor...

Страница 12: ...CK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use apen to manually disengage the locking mechanism...

Страница 13: ...2 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 220 Watts 220 Watts Centrifuge Motor H P Brushless H P Brushless Max G force 2 000 xg 1 600 xg 1 850 xg Max Speed...

Страница 14: ...e of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipm...

Отзывы: