Woodley Clinispin Horizon 6 Flex Скачать руководство пользователя страница 13

Woodley Equipment Company 

 +44 (0) 1204 669033 

 [email protected] | Page 10 

 

G

ENERAL 

S

PECIFICATIONS 

The rotor and accessories are rated for the maximum rotation frequency shown in the table below.

 

 

 

 

HORIZON 6 Flex

 

HORIZON 6 Flex FA

 

Tube Capacity 

6 tubes 

 3 to 10 mL

 

6 tubes 

 3 to 15 mL

 

Dimensions (H x W x D) 

23 cm x 30 cm x 36 cm  

(9 in x 12 in x 14 in)

 

23 cm x 30 cm x 36 cm  

(9 in x 12 in x 14 in )

 

Weight 

5.4 kg (12 lbs.)

 

5.4 kg (12 lbs.)

 

Sound Level 

64 dB A

 

64 dB A

 

Environmental Range 

16 

 32 degree C

 

16 

 32 degree C

 

Voltage 

95 -253 VAC

 

95 -253 VAC

 

Frequency 

50/60 Hz

 

50/60 Hz

 

Power Requirement 

220 Watts

 

220 Watts

 

Centrifuge Motor 

½ H.P. Brushless

 

½ H.P. Brushless

 

Max G-force 

2,000 xg

 

1,600 xg / 1,850 xg

 

Max Speed 

3,800 RPM

 

3,900 RPM

 

Cycle Time 

1 to 30 minutes (+/- 2%)

 

1 to 30 minutes (+/- 2%)

 

 
 
 

C

ALCULATING THE 

G-F

ORCE 

The I.F.U.s of tube manufacturers recommend cycles at a minimum G-force, which can be calculated if you know 
the RPM and the radius. Use the formula below.  

 

 

In Centimetres: 

RCF or G-force = 0.00001118 x 

Rotor Radius (cm) x (RPM)

2

 

In Inches: 

RCF or G-force = 0.0000284 x 

Rotor Radius (in) x (RPM)

2

 

 
 

 

 

HORIZON 6 Flex 

HORIZON 6 Flex FA 

 

 

Blue Tube (75 & 100 mm) 

Orange Tube (125 mm) 

Radius 

12.7cm (5 in) 

9.5cm (3.75 in) 

11cm (4.25 in) 

Содержание Clinispin Horizon 6 Flex

Страница 1: ...Clinispin HORIZON 6 Flex 6 Flex FA Operator s Manual...

Страница 2: ......

Страница 3: ...1 Initial Setup 2 Quick Start 3 Single Cycle Lock 4 Preset Lock 4 Setting or Modifying a Cycle 4 Settings 5 Operation 5 Balancing Loads 6 Care and Preventative Maintenance 7 Cleaning and Disinfection...

Страница 4: ...verheating of samples by using ambient air to keep specimens at room temperature o The tube holders are fibre reinforced for high strength durability and years of trouble freeuse o A clear lid permits...

Страница 5: ...and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be...

Страница 6: ...cycle displayed on the screen The lid must be closed Unlock Allows access into the rotor chamber by engaging the unlocking mechanism Entry is only possible when the rotor is stopped Stop Pressing the...

Страница 7: ...es can again be edited or programmed SETTING OR MODIFYING A CYCLE Up to 10 cycles can be named programmed for time speed and braking and saved in memory Cycle Selection Press the up and down buttons n...

Страница 8: ...ubes 1 850 xg N A 600 xg OPERATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise...

Страница 9: ...e life of the centrifuge and produce better results Use the following rules when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired...

Страница 10: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Страница 11: ...er with a tool to remove them o Inspect the rotor tube holders or buckets for damage o If the tube holders or buckets have any damage evenslight safely dispose of them and replace them o If the rotor...

Страница 12: ...CK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use apen to manually disengage the locking mechanism...

Страница 13: ...2 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 220 Watts 220 Watts Centrifuge Motor H P Brushless H P Brushless Max G force 2 000 xg 1 600 xg 1 850 xg Max Speed...

Страница 14: ...e of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipm...

Отзывы: