Cycle parameters cannot be changed

 

o

 

If cycle selection is locked on one cycle, press the UNLOCK 
button for 5 seconds. Then, press the GEAR button and follow 
the instructions on page 4 of this manual.

 

o

 

If different saved cycles can be selected but not modified, the 
centrifuge is in Preset Lock mode. Press the GEAR button, then 
the UNLOCK button for 5 seconds. You should now be able to 
change cycle parameters.

 

The centrifuge does not unlock after a 
run is completed

 

o

 

Wait until the rotor has come to a complete stop. If the lid knob 
still cannot be rotated, press the UNLOCK button and try again.

 

If no LED light is on, the unit is not 

powered and the lid will not unlock 
by conventional means. Remove the 
latch label and use a pen to manually 
disengage the locking mechanism. 
Pull the mechanism towards the 
control panel and then unlatch and 
open the lid.

 

o

 

If the unit is damaged, contact Technical Support for assistance.

 

The lid does not open

 

o

 

Ensure that the lid knob is turned fully counterclockwise.

 

o

 

If the knob cannot be turned counterclockwise, turn it fully 
clockwise, press UNLOCK, and turn counterclockwise.

 

o

 

If the lid remains locked after this and will not unlock, the 
electronics may have been damaged. Contact Technical 
Support for assistance.

 

Clicking noise during braking gets loud

 

Make sure that the screw in the centre of the rotor is tight.

 

Lid does not stay up

 

Tighten the centre screw on the lid hinge.

 

 

Содержание Clinispin Horizon 6 Flex

Страница 1: ...Clinispin HORIZON 6 Flex 6 Flex FA Operator s Manual...

Страница 2: ......

Страница 3: ...1 Initial Setup 2 Quick Start 3 Single Cycle Lock 4 Preset Lock 4 Setting or Modifying a Cycle 4 Settings 5 Operation 5 Balancing Loads 6 Care and Preventative Maintenance 7 Cleaning and Disinfection...

Страница 4: ...verheating of samples by using ambient air to keep specimens at room temperature o The tube holders are fibre reinforced for high strength durability and years of trouble freeuse o A clear lid permits...

Страница 5: ...and used in accordance with this operator manual may cause interference to radio communications Operation of this equipment in a residential area may cause interference in which case the user will be...

Страница 6: ...cycle displayed on the screen The lid must be closed Unlock Allows access into the rotor chamber by engaging the unlocking mechanism Entry is only possible when the rotor is stopped Stop Pressing the...

Страница 7: ...es can again be edited or programmed SETTING OR MODIFYING A CYCLE Up to 10 cycles can be named programmed for time speed and braking and saved in memory Cycle Selection Press the up and down buttons n...

Страница 8: ...ubes 1 850 xg N A 600 xg OPERATION o Place the tubes into the tube holders Be sure to follow the rules for balanced loads as listed in the next section o Close the lid and turn the lid knob clockwise...

Страница 9: ...e life of the centrifuge and produce better results Use the following rules when loading the rotor If an odd number of samples is to be spun fill a tube with water to match the weight of the unpaired...

Страница 10: ...the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is a spillage or...

Страница 11: ...er with a tool to remove them o Inspect the rotor tube holders or buckets for damage o If the tube holders or buckets have any damage evenslight safely dispose of them and replace them o If the rotor...

Страница 12: ...CK button and tryagain o If no LED light is on the unit is not powered and the lid will not unlock by conventional means Remove the latch label and use apen to manually disengage the locking mechanism...

Страница 13: ...2 degree C Voltage 95 253 VAC 95 253 VAC Frequency 50 60 Hz 50 60 Hz Power Requirement 220 Watts 220 Watts Centrifuge Motor H P Brushless H P Brushless Max G force 2 000 xg 1 600 xg 1 850 xg Max Speed...

Страница 14: ...e of their waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment The separate collection and recycling of your waste equipm...

Отзывы: