background image

6

MAINS CONNECTION 

Refrigeration equipment is commonly used in 

environments located close to liquids, and is 

usually operated in and around damp areas, 

or where restricted movement for installation 

and service is evident.

Great care must be exercised at all times 

when installing, operating, or servicing this 

appliance.

The equipment is to be powered by a suitable 

socket outlet which provides means of switching 

off and/or disconnecting the power supply to the 

equipment for cleaning or maintenance purposes.  It 

should be positioned in a clearly visible location, and 

be easily accessible at all times.  It is advised that 

a label be fixed adjacent to the designated socket 

outlet for identification purposes, stating the outlet 

is used to power the specific medical refrigeration 

equipment.

For appliances fitted with a moulded plug for safety, 

ensure that the mains power cable is extended 

free from the refrigeration system to avoid 

entanglement. If a plug or mains cable requires 

replacement, contact the Williams Spares Office on 

+44(0)1553 817017.

The installation of a fixed appliance and periodic 

inspection should only be undertaken by a qualified, 

skilled, and competent electrician; and connected 

to the correct power supply suitable for the load as 

stipulated by the appliance data label. 
The electrical installation and connections should 

meet the necessary requirements to the local 

electrical wiring regulations and any electrical safety 

guidelines. 

All appliances rely upon a suitable connection 

to earth to ensure safe operation. If in doubt, 

contact a qualified, skilled, and competent 

electrician before using the appliance.

We recommend:-
• Supplementary electrical protection with the  

  use of a residual current device (RCD)
• Fixed wiring appliances incorporate a switch  

  disconnector for purposes of safe isolation        

  to meet the specification requirements of IEC      

  60947

SUPERVISORY & NETWORK CONNECTIVITY 

(OPTIONAL) 

There is an option to connect the Medi+ 

refrigeration equipment to third party supervisory 

networks and systems such as Building 

Management Systems (BMS), and IoT cloud based 

systems via an RS485 Modbus data connection.  This 

will allow for a comprehensive supervision of the 

equipment operating parameters and status, such 

as internal temperature, defrost cycles and alarms.
The Medi+ equipment controller also provides a volt 

free contact, which can be connected to a BMS to 

monitor general alarm conditions.

NB:  A proprietary cable from Williams 

Refrigeration is required and, only suitably 

qualified and competent engineers should 

install this equipment to such systems.  For 

more information, please contact Williams 

Refrigeration on +44(0)1553 817000.

 

 

MAINS POWER SUPPLY FAILURE ALARM 

(OPTIONAL) 

This alarm is to help safeguard stored medical 

samples/items by announcing when a failure of the 

mains power supply has occurred.  An audible (95dB 

sounder), and a visual alarm (red flashing ‘Power 

failure’ indicator) will activate in the advent of a 

mains power supply failure.
The alarm will be continuous for a minimum of 

12 hours, (whilst the mains power supply failure 

remains), powered by a rechargeable battery 

integrated into the alarm circuit; however, the 

sounder can be muted at any time by pressing the 

‘Mute’ button. Once muted, a blue flashing indicator 

signifies that the alarm has been muted. The red 

flashing ‘Power failure’ indicator will continue to be 

active whilst there is a failure of the mains power 

supply.
When the mains power supply is reinstated the 

‘Alarm muted’ function MUST be reset by pressing 

the mute button once more, cancelling the blue 

flashing indicator signifying that the alarm circuit is 

ready to sense a power failure.

IMPORTANT: for the rechargeable battery 

to receive a full charge, it is advised that 

the mains power supply is connected for 

at least 24 hours. As with all batteries, the 

rechargeable battery has a finite operational 

life which will vary dependent of a number of 

factors, it is therefore, advised that the mains 

power supply failure alarm is checked by a 

service engineer as part of the routine periodic 

servicing of the refrigerator. Only replace the 

battery with a Williams Refrigeration approved 

and supplied battery.

INSTALLATION

Содержание Medi+ LWMP620

Страница 1: ...nd will use the necessary safety equipment and take the necessary precautions required of their intended work Improper installation maintenance or repair may put the user at considerable risk Williams...

Страница 2: ...led by persons who are appropriately trained skilled and competent to properly and safely carry out the work and serviced by qualified engineers for the handling of hydrocarbon refrigerants Ensure pro...

Страница 3: ...the replacement of parts is required the equipment must be safely disconnected and isolated from the power supply using a lock off system at the isolation device It is advised that the electrical sup...

Страница 4: ...iosafety Manual source World Health Organisation Laboratory Biosafety Manual the current edition IMPORTANT SAFETYINFORMATION INFORMATION ON PRODUCT LIABILITY In the following cases the designated prot...

Страница 5: ...ay to reveal the polished stainless steel surface Care should be taken to ensure that no adhesive residue remains on the surface Any stubborn or tough adhesive marks can be removed by following the ad...

Страница 6: ...party supervisory networks and systems such as Building Management Systems BMS and IoT cloud based systems via an RS485 Modbus data connection This will allow for a comprehensive supervision of the eq...

Страница 7: ...e even loading Do not place items against the rear plenum or pushed against side walls When loading the appliance please ensure that its load is equally distributed throughout and ensure air can circu...

Страница 8: ...essing the i button While keeping the i button pressed press the x button to reset the value To exit the information menu either wait 10 seconds or press the x button Alarms An alarm is reported via t...

Страница 9: ...tinue in normal sequence Instead the compressor will switch to a five minute cycle which it cycles on for five minutes then off for five minutes The normal compressor cycle will be restored upon the r...

Страница 10: ...ugher spots creamy polishes like CIF original cream can be effective Light pressure should be used when cleaning with the grain The cleaning process should be repeated in order to prevent any dirt bec...

Страница 11: ...hould be checked and cleaned regularly and replaced if damaged To clean the gasket wipe with warm soapy water and a soft cloth ensuring it is completely dry before closing the door DO NOT use a sharp...

Страница 12: ...ing Shut door Remove cover and clean condenser fins with a clean brush TROUBLE SHOOTING INFORMATION FOR THE MAINS POWER SUPPLY FAILURE ALARM OPTIONAL Fault Possible Cause Action The sounder is not wor...

Страница 13: ...same quality and standard as we use to build your appliance and have been rigorously checked tested and inspected to ensure the very best quality and exact fit You can contact us directly for everythi...

Страница 14: ...number and date of installation The serial number is located on the product identification plate inside the cabinet modular product door frame or similar location It is recommended that operators shou...

Страница 15: ...ent that may affect performance 14 The customer permits persons other than those authorised by Williams Refrigeration to perform or affect repairs or adjustments to the equipment 15 If authorised repr...

Страница 16: ...k Williams Refrigeration Hong Kong Unit C 12 F Roxy Industrial Centre 58 66 Tai Lin Pai Road Kwai Chung New Territories Hong Kong Tel 852 2407 5422 Fax 852 2407 3767 sales williams hongkong com www wi...

Отзывы: