6 Adjustment with the display and adjustment module
WIKA Operating Instructions - Differential pressure transmitter DPT-10
Min. adjustment
0.00 %
=
0.0000 bar
0.0000 bar
2. Set the requested value with
[+]
and
[->]
, for example 100 %.
3. Confirm with
[OK]
and edit the requested bar value.
4. Set the requested bar value with
[+]
and
[->]
, for example
29.4 mbar.
5. Confirm with
[OK]
and move to max. adjustment with
[->]
.
For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.
The min. adjustment is finished.
For the max. adjustment with density, a filling of the vessel is not
necessary. The numeric examples are stated in chapter
Mounting
,
Measurement setup, Density and interface
of this instructions manual.
Proceed as follows:
1. Edit the % value in the menu item "
Max. adjustment
" with
[OK]
.
100.00 %
=
0.1000 bar
0.0000 bar
Information:
If the instrument has not yet been adjusted, the displayed pressure for
100 % corresponds to the nominal measuring range of the sensor (in
the above example 100 mbar).
2. Set the requested value with
[->]
and
[OK]
, for example 0.0 %.
3. Confirm with
[OK]
and edit the requested mbar value.
4. Set the requested value with
[+]
and
[->]
, for example 35.3 mbar.
5. Confirm with
[OK]
and move to the menu overview with
[ESC]
.
For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.
The max. adjustment is finished.
Proceed as follows:
1. Edit the bar value in the menu item "
Min. adjustment
" with
[OK]
.
Min. adjustment
0.00 %
DP
- +
=
0.0000 bar
0.0000 bar
2. Set the requested bar value with
[+]
and
[->]
.
3. Confirm with
[+]
and move to max. adjustment with
[->]
.
For an adjustment with flow, simply enter the actual measured value
indicated at the bottom of the display.
Basic adjustment - Max.
adjustment with density
Basic adjustment - Min.
adjustment with flow