72
11 Supplement
WIKA Operating Instructions - Differential pressure transmitter DPT-10
Measuring cell
10 mbar
30 mbar
100 mbar
500 mbar
Influence of the sys-
tem pressure to the
zero point
±0.15 % of
URL/7 bar
±0.50 % of
URL/70 bar
±0.15 % of
URL/70 bar
±0.075 % of
URL/70 bar
Influence of the sys-
tem pressure to the
span
±0.035 % of
URL/7 bar
±0.14 % of
URL/70 bar
±0.14 % of
URL/70 bar
±0.14 % of
URL/70 bar
Measuring cell
3 bar
16 bar
40 bar
Influence of the system
pressure to the zero point
±0.075 % of URL/7 bar
±0.075 % of URL/70 bar
±0.075 % of URL/70 bar
Influence of the system
pressure to the span
±0.14 % of URL/7 bar
±0.14 % of URL/70 bar
±0.14 % of URL/70 bar
Alloy gold rhodium-coated diaphragm
Measuring cell
10 mbar
30 mbar
100 mbar
500 mbar
Influence of the sys-
tem pressure to the
zero point
±0.15 % of
URL/7 bar
±0.77 % of
URL/70 bar
±0.42 % of
URL/70 bar
±0.075 % of
URL/70 bar
Influence of the sys-
tem pressure to the
span
±0.035 % of
URL/7 bar
±0.14 % of
URL/70 bar
±0.42 % of
URL/70 bar
±0.14 % of
URL/70 bar
Measuring cell
3 bar
16 bar
40 bar
Influence of the system
pressure to the zero point
±0.075 % of URL/7 bar
±0.075 % of URL/70 bar
±0.075 % of URL/70 bar
Influence of the system
pressure to the span
±0.14 % of URL/7 bar
±0.14 % of URL/70 bar
±0.14 % of URL/70 bar
Tantalum diaphragm
Measuring cell
10 mbar
30 mbar
100 mbar
500 mbar
Influence of the sys-
tem pressure to the
zero point
±0.32 % of
URL/7 bar
±1.60 % of
URL/70 bar
±0.42 % of
URL/70 bar
±0.14 % of
URL/70 bar
Influence of the sys-
tem pressure to the
span
±0.07 % of
URL/7 bar
±0.32 % of
URL/70 bar
±0.42 % of
URL/70 bar
±0.14 % of
URL/70 bar
Measuring cell
3 bar
16 bar
40 bar
Influence of the system
pressure to the zero point
±0.14 % of URL/7 bar
±0.14 % of URL/70 bar
±0.14 % of URL/70 bar
Influence of the system
pressure to the span
±0.14 % of URL/7 bar
±0.14 % of URL/70 bar
±0.14 % of URL/70 bar