Problem
Possible cause
Solution
The PV display
keeps
indicating
the value set in
[Scaling low limit].
Check whether the input signal
wires of DC voltage (0 to 5 V
DC, 0 to 10 V DC) and direct
current (0 to 20 mA DC) is
disconnected.
How to check whether the input signal wire is
disconnected
[DC voltage (0 to 5 V DC, 0 to 10 V DC)]
If the input to the input terminal of this instrument is 1
V DC, and if a value corresponding to 1 V DC is
indicated, the instrument is likely to be operating
normally, however, the input signal wire may be
disconnected.
[Direct current (0 to 20 mA DC)]
If the input to the input terminal of this instrument is 4
mA DC, and if a value (converted value from scaling
high, low limit setting) corresponding to 4 mA DC is
indicated, the instrument is likely to be operating
normally, however, the input signal wire may be
disconnected.
Check whether the input
terminals of DC voltage (0 to
5 V DC, 0 to 10 V DC) or direct
current (0 to 20 mA DC) are
securely connected to the
instrument input terminals.
Connect the input terminals of DC voltage and current
to the input terminals of this instrument securely.
The indication of
PV
display
is
irregular
or
unstable.
Check whether sensor input or
temperature unit ( or
) is
correct.
Select the sensor input and temperature unit ( or )
correctly.
Sensor correction coefficient or
Sensor correction value is
unsuitable.
Set them to suitable values.
Check whether the sensor
specification is correct.
Use a sensor with appropriate specifications.
AC leaks into the sensor circuit. Use an ungrounded type sensor.
There may be equipment that
interferes with or makes noise
near the instrument.
Keep the instrument clear of any potentially disruptive
equipment.
Values on the PV
display do not
change.
Terminals 14 and 17 are
closed, and the HOLD function
is working.
Cancel the HOLD function by opening terminals 14 and
17.
Even if the UP key
or DOWN key is
pressed, values do
not change.
Set value lock (Lock 1 or Lock
2) is selected.
Release the lock in [Set value lock].