WIKA operating instructions premium pressure transducer, model CPT9000
0020504001G 10/2019 EN
EN
9.2 General Specification
Case
Orientation effects
Negligible – completely removable with re-zeroing
Dimensions
See technical drawings
Weight
~250 g (depending on range)
Display
Resolution
100 ppb or better
Warm-up time
15 min.
Connections
Pressure connections
SAE J514/JIC 4
Overpressure limit
2X proof, 3X burst; static pressure < 50 psig
Materials, wetted parts
Ranges ≤ 5 psi
– Silicon, 316 SS, glass filled resins, epoxy
Ranges > 5 to 1500 psi – 316 SS
Ranges >1500 psi – 316 SS, fluorocarbon rubber
Pressure media
Ranges
≤
5 psi – clean, dry, non-corrosive gases
Ranges > 5 psi – media compatible with the listed wetted parts
Voltage supply
Power supply
9 to 18 VDC (12 VDC nominal)
Power consumption
< 26 mA at 12 VDC +/-5%
Permissible ambient conditions
Compensated temperature range 0 to 50
o
C (32 to 155
o
F)
Operating temperature range
-40 to 85
o
C (-40 to 185
o
F)
Storage temperature range
-40 to 85
o
C (-40 to 185
o
F)
Humidity
0 ... 95% r.h. (non-condensing)
Operating altitude
<3000 meters (10,000 feet)
Operating environment
IP-67
Measure port internal volume
<1 cc
Reference port internal volume
~ 40 cc
Communication
Interface
RS-232 or RS-485
Baud rate
Default 57,600 baud - 9600, 19200 and 115200 user selectable
Measuring rate
50 values/second, default - (factory adjustable)
Command sets
See manual, Section 6 Operation
9.3 Approvals and Certificates
Compliance
CISPR 11 & A1, IEC 61000-4-2, 3, 4, 5, 6, 11, IEC 61000 A1, A2, IEC 61010-1
Calibration
4)
A2LA calibration certificate
4) Calibration in a vertical position.
Specifications