WIKA operating instructions premium pressure transducer, model CPT9000
0020504001G 10/2019 EN
EN
6.1.2.4
Limits for Pressure Alarms Commands
PRESS_LIM_MIN & PRESS_LIM_MAX – These commands set the limits for pressure alarms. If the pressure exceeds the
PRESS_LIM_MAX, an error will be generated and placed in the error queue. Similarly, if the pressure drops below the PRESS_
LIM_MIN, an error will be generated and placed in the error queue.
6.1.2.5
Setting Pressure Units
UNIT_INDEX - This command sets the output units per the following table:
Unit Index Units
Unit Index
Units
1
psi
21
Torr
2
inHg 0C
22
kPa
3
inHg 60F
23
Pa
4
inH2O 4C
24
dy/cm2
5
inH2O 20C
25
g/cm2
6
inH2O 60F
26
kg/cm2
7
ftH2O 4C
27
MSW 0C
8
ftH2O 20C
28
osi
9
ftH2O 60F
29
psf
10
mTorr
30
tsf
11
inSW 0C
31
Not Used
12
ftSW 0C
32
uHg 0C
13
atm
33
tsi
14
bar
34
mHg 0C
15
mbar
35
hPa
16
mmH2O 4C
36
Mpa
17
cmH2O 4C
37
mmH2O 20C
18
MH2O 4C
38
cmH2O 20C
19
mmHg 0C
39
mH2O 20C
20
cmHg 0C
99
CUST_UNIT
Table 6.1.2.5 - Available Unit Codes and Units
6.1.2.6
Customer Units
The CUST_UNIT multiplier in conjunction with UNIT_INDEX 99 allows the user to set their own units. The CUST_UNIT is a
multiplier times psi to achieve the customer’s unit value.
6.1.2.7
Calibration Adjustments
SPAN & ZERO - These are calibration multipliers and offset respectively. The zero is an offset applied to all output pressure
values. The span is a multiplier times all pressure output values.
6.1.2.8
Temperature Limits and alarms
TEMP_LIM_MIN & TEMP_LIM_MAX – These set the temperature limits that trip alarms just like the PRESS_LIM_MIN and MAX.
6.1.2.9
Sea level and Altitude Units Settings
SEA_LEVEL & ALT_UNIT – These commands set the altitude units in meters or feet and allow the user to set a sea level
adjustment in altitude units. This is only applicable for absolute pressure transducers with a maximum range of <= 30 psia.