WIKA operating instructions premium pressure transducer, model CPT9000
0020504001G 10/2019 EN
EN
Operation
N/A
PWD_CHANGE<sp><n,n>
nnnn,nnnn
Change password <n,n>
= <old,new>
N/A
PWD<sp><n>
nnnn
Send password
Pressure Related Commands
PRESS?
N/A
Returns configurable
output from active
channel in current units
see Table 6.1.2.10
& Output Control
Commands section.
RANGE_MIN?
N/A
+n.nnn
Minimum range in
current units.
RANGE_MAX?
N/A
+n.nnn
Maximum range in
current units
PRESS_LIM_MIN?
PRESS_LIM_MIN<sp><n>
+n.nnn
Low alarm limit in current
units. See 6.1.2.4
PRESS_LIM_MAX?
PRESS_LIM_MAX<sp><n>
+n.nnn
High alarm limit in
current units. See
6.1.2.4
UNIT_INDEX?
UNIT_INDEX<sp><n>
n
Mensor unit index: 1-39
or 99 for CUST_UNIT.
See Table 6.1.2.5
UNIT?
N/A
cccccccccc
Unit text 10 char -
mmHg, psi, kPa…
CUST_UNIT?
CUST_UNIT<sp><n>
+n.nnn
Sets custom unit
multiplication (xx/psi).
See Section 6.1.2.6
Calibration Related Commands
CAL_DATE?
CAL_DATE<sp><yy,mm,dd>
(Password Protected)
nn,nn,nn
Calibration date
SPAN?
CAL_SPAN<sp><n>
(Password Protected)
+n.nnn
Customer span multipler
- 0.99 to 1.01. See
Section 6.1.2.7
ZERO?
CAL_ZERO<sp><n>
(Password Protected)
+n.nnn
Customer zero adder.
See Section 6.1.2.7
INTERVAL?
CAL_INTERVAL<sp><n>
(Password Protected)
n
Calibraion interval in
days
TARE?
TARE<sp><n>
n
Turns tare on/off - 1/0
TARE_OFFSET?
N/A
+n.nnn
Gets the offset if Tare is
set to 1
Linearity Related Commands
LIN_COEFF?
CAL_LIN_COEFF<sp><n,n>
(Password Protected)
+n.nnn
Customer linearity
coefficients
LIN_SEGS?
CAL_LIN_SEGS<sp><n>
(Password Protected)
n
Linearity correction
segments 0 to 10
Temperature Related Commands
TEMP?
N/A
+nnn.n
Temperature in deg. C.