29
WIKA operating instructions premium pressure transducer, model CPT9000
0020504001G 10/2019 EN
EN
7.2.6 Calibration & Communication Software
The Operation and Calibration software utility is available on the Mensor.com website from the main menu select
Service/
Calibration
then
Download Software
then
PC Software
.
The purpose of this software is to communicate with the CPT9000 transducer. Calibration of the transducer output and
configuration of various parameters can be accomplished. It has many features that allow you to read and adjust the rich feature
set including:
• Changing filter settings
• Modifying custom identification strings
• Showing the CPT9000’s error stack
• Setting the tare pressure on the CPT9000
• Modifying the output format of the sensor
• Changing the pressure units output
It also has many features to adjust various calibration parameters after the appropriate password is sent:
• Setting the calibration date
• Setting the calibration interval
• Setting custom pressure units
• Setting sea level adjustment on barometric range transducers
• Setting minimum and maximum pressure alarm limits
• Setting minimum and maximum temperature alarm limits
• Setting Zero and Span
• Adjusting the CPT9000 linearity
• Resetting the calibration password
• Restoring the factory calibration