WIKA operating instructions, model CPH7650
EN
14261987.02 01/2019 EN/DE
6. Commissioning, operation
6.6.6 CALIBRATION mode (calibration of a pressure gauge)
1st test point
(define e.g. 0 bar)
123456
ID no:
02
P-01:
bar
0.000
True:
bar
1.000
Set:
%FS
-0.10
bar
-0.006
Dev.:
bar
1.000
6.000
0.000
Testitem
bar
1.006
10.000
0.000
Reference
13:41
KALIBRIEREN
123456
ID no:
02
P-01:
bar
0.000
True:
bar
1.000
Set:
%FS
0.00
bar
0.000
Dev.:
bar
0.000
6.000
0.000
Testitem
bar
0.000
10.000
0.000
Reference
13:41
Calibration
123456
ID no:
01
P-01:
bar
0.000
True:
bar
0.000
Set:
%FS
0.00
bar
0.000
Dev.:
bar
0.000
6.000
0.000
Testitem
bar
0.000
10.000
0.000
Reference
13:41
Calibration
Deviation
Deviation
Deviation
No. of test item and
No. of test point
No. of test item and
No. of test point
No. of test item and
No. of test point
Generate the specified set
point in accordance with the
test item display using the
pressure generator
If the set point = 0, make sure
the calibration assembly is
in a pressure-free condition/
vented to atmosphere (test
item must indicate 0 bar; zero
point setting if necessary)
and, with
[ENTER]
, record
the measured values of this
test point
Generate the specified set
point in accordance
with the
reference display
using the
pressure generator
Record the measured values
of the test point
(with
[BACK]
return to the
previous test point)
ENTER
ENTER
2nd test point
(define e.g. 1 bar)
2nd test point
(defined)
xth test point
(define)