WIKA operating instructions, model CPH7650
EN
14261987.02 01/2019 EN/DE
4. Design and function
4.5 Connection of the test item
Only use original parts (see chapter 11 “Accessories”).
Above the electrical connections is located the mechanical pressure connection for the
connection of a test item. This is marked on the front of the calibrator with “Test Port”.
1
Connection of the test item
Mechanical and also electrical pressure measuring instruments can be connected. For
the connection of the electrical connections, see chapter 4.3 “Electrical connections to
▶
The instrument to be tested is inserted into the quick-release connector with knurled
nut and can be oriented.
⇒
A hand-tight fastening is sufficient for a secure sealing.
▶
The O-ring seal on the test connection must be seated correctly and not worn. Replace
them if necessary.
▶
It is important that any instrument that is to be connected is clean.
⇒
To prevent contamination by the test items, the use of a dirt trap is recommended!
■
The quick-release connector is fitted with a G 1/2 threaded insert as standard.
⇒
For further threaded inserts and adapters, see chapter 11 “Accessories”.
For short threads, a sealing insert may be required.
- 24V OUT +
- V +
- mA / +
V
max. 12V
-
+
sig+
max. 24mA
mA
-
+
sig-
sig+
sig+
sig+
24V
output
-
+
UB-
UB-
sig+
UB+
UB+
self-powered
no
no
no
yes
test item
2wire (mA)
2wire (mA)
3wire (mA)
3wire (V)
1
.
2
5
7
9
+/-
CLEAR
ENTER
0
8
6
4
3
BACK
SETUP
+ Pessure -
Fine Adjustment
Vent
Reference
Sensor
Reference
Sensor
Connection
Test Port
CPH7650
1