7  Maintenance 

7.1  Adjusting 

Menu item 12 in UNIGAS 300 can be used to adjust the temperature sensor and/or the pressure sensor to 
increase the measurement accuracy. 

The pressure sensor can be adjusted with an offset value (p_offset). Modifying this value will result in that the 
measured pressure over the entire range is increased or decreased by this value. 

The temperature sensor can be adjusted by an offset value (tcorr_0) and by entering a span. This span can 
be set for the measuring range from -40 °C to 0 °C (tcorr_min.) and for the measuring range from 0 to 55 °C 
(tcorr_max). 

As a rule, tcorr_0 is set at -0.24 °C for correction of the 3-m length of cable with which the temperature sensor 
is connected. 

The value of span is expressed in °C/°C and can be determined as follows (X °C is a random temperature 
higher than 0 °C or a random temperature below 0 °C): 

 

 

 

 

 

Span = deviation at X °C – deviation at 0 °C 

 

 

 

 

 

 

 

X °C – 0 °C 

 

Proceed as follows: 

 

12. Adjusting 

 

Screen 12-x: p_offset, tcorr_min, tcorr_0, tcorr_max 

 

for activating, value starts blinking 

 

 

 

and 

 

 to modify the value 

 

 

 

confirm, confirmation will only be accepted if the 

programming switch is operated simultaneously. The 
value will stop blinking. 

 

 

 

   to return  

 

7.2  Replacing the pressure sensor or the temperature sensor 

The seal must be broken in order to replace the pressure sensor or the temperature sensor. Then unscrew the 
bolt, see chapter 4, fig. 4, item 5. 

Now remove the battery, the cables to the terminal strip and the protective cover. 

Replace the pressure sensor or the temperature sensor and put the protective cover and the cables back in 
place. Finally, put the battery back. When the housing is closed, UNIGAS 300 will automatically take over the 
data from the pressure sensor. This can be checked by the serial number, see menu item 9 (System). 

Subsequently, UNIGAS 300 must be checked for accuracy. As the power to UNIGAS 300 has been interrupted, 
it must be checked whether the clock is running correctly. If necessary, the clock must be set.  

 

Содержание UNIGAS 300

Страница 1: ...Installation operating and maintenance manual UNIGAS 300...

Страница 2: ......

Страница 3: ...ecked and corrected where applicable This particularly applies where safety aspects must be taken into account Further support can be obtained from the branch or representative responsible for your ar...

Страница 4: ...ions 12 3 3 1 Impulse outputs 12 3 3 2 Alarm outputs 12 3 3 3 Battery consumption counter 12 4 Installation 13 4 1 Main components 13 4 2 Connections 15 4 3 Connecting a gas meter with encoder output...

Страница 5: ...ts Namur input for gas meters with HF impulse output or encoder output two alarm inputs for normally close contacts two programmable real time impulse outputs two alarm outputs programming using softw...

Страница 6: ...ionyl chloride DD cell G8610080000 Test valve type BDA 04 for connecting pressure calibration equipment O31300 Software UNITOOL G6900000 Universal fastening set including mounting material GG6390 Flan...

Страница 7: ...ating and maintenance manual UNIGAS 300 DDG6004MHGB 08 2017 Rev A2 Software version from M 1 3 40 and D 2 3 37 5 2 Explosion safety instructions Ex See DDG6800CVML EU Declaration of Conformity and Saf...

Страница 8: ...oval metrology MID certificate number 3 MID and conversion functions 3 1 Measurement Instruments Directive MID UNIGAS 300 is MID approved and complies with standard EN12405 1 A2 Calibration relevant d...

Страница 9: ...1 HF Cf1 Qf1 Qmin gas meter Cf2 Qf2 Cf9 Qf9 Cf10 Qf10 Qmax gas meter T p p meter T meter Gas meter LF HF Encoder Gasmeter status bits 1 7 NAMUR_short_circuit 1 8 NAMUR_interruption INP1DIV Encoder de...

Страница 10: ...ible for conversion CRC error conversion counting will continue in Vc1err conversion takes place in Vb1err with relevant replacement values o pressure measurement fails or is outside range of pmin to...

Страница 11: ...es with standard EN12405 1 A2 The corrected volume is counted in counter Vc1 This will be discussed further in chapter 12 If a low frequency impulse or encoder input is used the function of error corr...

Страница 12: ...tware version D 2 3 37 the function has been added that allows channel 6 Vm3 of the interval logger to be set for logging of Hs values see also table Gas Composition and Redirection on page 39 When re...

Страница 13: ...ith a date time stamp Characteristics 360 items ring memory readable through serial communication Stored items include date time bit number and status register number including the information regardi...

Страница 14: ...oupled to a status bit from one of the three status registers At the moment when the status bit is created an impulse of 0 1 s is issued As long as the status bit is active a 0 1 s impulse is issued e...

Страница 15: ...l see chapter 3 2 Housing lock Unscrew these two bolts until the bolt heads are roughly level with the UNIGAS 300 front Then open the housing on the right hand side The seal may stick so it may take s...

Страница 16: ...sma Sikkema for more information 11 Dot matrix display The display is activated when one of the navigation keys is operated If the navigation keys are not operated for 1 min the display will be switch...

Страница 17: ...which means that manipulation of the module connector will not affect the calibration functions 8 Tamper switch UNIGAS 300 detects the opening of the housing and records this event in status register...

Страница 18: ...r supply with galvanic separation is strongly recommended Outputs 13 ALARM OUT 2 alarm output 2 14 GND ALARM common earth for ALARM OUT 1 and 2 15 ALARM OUT 1 alarm output 1 16 PULSE OUT 2 impulse out...

Страница 19: ...ust be shielded UNIGAS 300 comes with cable glands fitted with EMC shielding to which the shield is to be connected according the step by step mounting instruction below 1 Remove the sheath over 35 mm...

Страница 20: ...Software version from M 1 3 40 and D 2 3 37 18 7 Press the cable into the cable gland until the shield is in the cable gland 8 Press the seal into the cable gland and tighten it slightly 9 Tighten th...

Страница 21: ...the gas meter or at a different position in the gas installation that meets the following criteria distance to the gas meter not exceeding 1 m position downstream of the turbine gas meters insertion...

Страница 22: ...pressure measuring line 2 1 More info Press key UNIGAS 300 Figure 6 Mounting of BDA 04 with special connection adapter Mount a BDA 04 as follows 1 place the BDA 04 1 next to the pressure sensor 2 fast...

Страница 23: ...igure 7 Calibration lock and optional housing seal Figure 7 shows the positions for the seals 1 adhesive seal for calibration lock 2 adhesive seal for access to bottom PCB 3 adhesive seal for access t...

Страница 24: ...l indicates a partly empty battery Check the value in UNIGAS 300 menu item 10 When the remaining capacity drops below 10 the battery condition indicator will start blinking indicating that the battery...

Страница 25: ...M 1 3 40 and D 2 3 37 23 Menu screen Figure 9 Elements of the menu screen 1 selected menu item 2 menu title 3 number of selected menu item 4 indicator that more screens can be displayed if or is actua...

Страница 26: ...register An alarm cannot be reset until the cause of that alarm no longer exists 8 2 x Operation status register 2 8 3 x Alarm status register 3 See chapter 10 for more information Remark if there ar...

Страница 27: ...ers is represented in a comparable way but pre leading 0 are not shown On the pages that follow the three status registers are explained in more detail The tables successively represent the name with...

Страница 28: ...ODER 2 3 6 7 A B E F C 9 Readout values from encoder counter are unusable BCC error or a encoder reading is lower than counter reading Vm1 or Vm2 Vm1 or Vm1 will not be modified NAMUR open circuit 1 3...

Страница 29: ...an value Vm1Vm2_warning See chapter 17 for more details Alarm tmax 22 8 9 A B C D E F C B Temperature Alarm tmax Alarm tmin 4 5 6 7 C D E F C A Temperature Alarm tmin Alarm pmax 2 3 6 7 A B E F C 9 Pr...

Страница 30: ...s full the oldest loggings will be overwritten Is cancelled as soon as log book is deleted Alarm input 2 2 3 6 7 A B E F C 9 Connected alarm contact open Alarm input 1 1 3 5 7 9 B D F C 8 Connected al...

Страница 31: ...or a random temperature below 0 C Span deviation at X C deviation at 0 C X C 0 C Proceed as follows 12 Adjusting Screen 12 x p_offset tcorr_min tcorr_0 tcorr_max for activating value starts blinking...

Страница 32: ...ensor tcorr_0 must be set according following table to correct the length of the cable of the temperature sensor Length cable tcorr_0 1 5 m 0 05 C 3 m 0 1 C 5 m 0 17 C 10 m 0 33 C Replacing the intern...

Страница 33: ...he replacement pressure sensor in the reverse order When exchanging an external pressure sensor the shield of the pressure sensor cable must be connected with the bush in the cable gland see chapter 4...

Страница 34: ...al voltage 3 6 V nominal capacity 35 Ah initial capacity to be entered into UNIGAS 300 21 Ah applicable in Western Europe Observe the procedure below before the battery is replaced Check that the pack...

Страница 35: ...performed when the housing is opened 10 Battery Screen 10 1 Status Screen 10 1 1 Status U batt AH used AH new Screen 10 2 Replace Screen 10 2 1 Initial capacity Screen 10 2 1 1 Ah for activating value...

Страница 36: ...Installation operating and maintenance manual UNIGAS 300 DDG6004MHGB 08 2017 Rev A2 Software version from M 1 3 40 and D 2 3 37 34 9 Sensor pockets for temperature sensor Available models...

Страница 37: ...Password 2b is protected by password 2b Protection level 2c With password 2c it is possible to adjust the gas composition CO2 H2 N2 d and Hs Password 2c is protected by password 2c If no passwords hav...

Страница 38: ...alue based upon 5 minutes interval Qb1_nx5 m3 h 3 2 7 1 43 2 2 Flow at base conditions undisturbed value based upon the moving average of n x 5 minutes interval Qb1_inst m3 h 3 2 7 1 43 2 0 Instantane...

Страница 39: ...for measuring error correction of gas meter Between Qf3 and Qf4 linear interpolation takes place between Cf3 and Cf4 Qf3 m3 h C 98 6 Qf3 where Qf2 Qf3 Qf4 Cf4 C 98 7 Factor 4 for measuring error corr...

Страница 40: ...range within which conversion takes place tmin C 2 3 C 97 8 Minimum temperature of the temperature range within which conversion takes place tmax C 2 3 C 97 9 Maximum temperature of the temperature ra...

Страница 41: ...ation lock on factory set o calibration lock off o calibration lock off on condition that the calibration log book is not yet full contents overwriting Z and Zb calculation 1 1 C 93 28 When the calcul...

Страница 42: ...aximum value of Qc_nx5 last year Qc_nx5 current year m3 h C 95 56 Maximum value of Qb_nx5 last year Settings for alarm messages and warnings Protection level 1 password 2b Name Unit OBIS code Descript...

Страница 43: ...of UNIGAS 300 Ah_used Ah 10 1 1 C 6 1 Battery capacity used by UNIGAS 300 This value will be reset when the battery is replaced through menu item 10 2 Ah_new Ah 10 1 1 C 6 4 Available capacity batter...

Страница 44: ...quals 00000000 the device will respond to any device address or to no device address Device_adress_61E C 90 26 Device address on which UNIGAS 300 will respond as a UNIGAS 61 E Presentation display clo...

Страница 45: ...larm output 2 C 93 5 Selector switch for relaying a status bit to alarm output 2 If the status is active and persistent an impulse will be sent to the output every 5 min interval at the moment when it...

Страница 46: ...Approval MID 002 MID 2004 22 EC by NMi NMi T10132 Measuring interval for pressure and temperature 5 25 s configurable standard 25 s Conversion interval LF every impulse HF every second if impulses are...

Страница 47: ...once a day and 15 min display use per month PTZ version typically 15 years without use of encoder input typically 5 years with use of encoder input 25 s measuring interval typically 13 years with use...

Страница 48: ...wer necessary 1000 h emergency power supply from D cell in case of power failure Maximum frequency 5 kHz Minimum impulse duration 100 s Error curve correction Linear interpolation on the basis of a ma...

Страница 49: ...NX19 modified Gasunie SGERG TM5 1991 method 1 4 AGA 8 gross method 1 Population Complete algorithm implemented in UNIGAS 300 Data storage Loggers interval logger 5 min interval 150 days 43200 5 min re...

Страница 50: ...ction Screw terminals Specification of switching voltage and current Suitable for switching voltage of 3 20 V and switching current of 6 uA 50 mA Communication ports Communication port 1 for modem ser...

Страница 51: ...m3 h Exceeding of alarm Vc1_60 Corrected unconverted consumption in clock hour configurable between 1 and 10 000 m3 Exceeding of warning Vc1_60 Corrected unconverted consumption in clock hour configur...

Страница 52: ...01 Explosion hazard EN IEC 60079 0 Explosive atmospheres Part 0 General requirements 2012 EN IEC 60079 11 Explosive atmospheres Part 11 Equipment Protection by Intrinsic Safety 2012 EN IEC 60079 28 Ex...

Страница 53: ...error must be converted into the connected Cf value for each point Cf ____100____ measuring error 100 Where Measuring error listed gas meter measuring error expressed in Example a measuring error of 2...

Страница 54: ...erence Vm2 Vm1 If the volume difference exceeds the preset alarm value UNIGAS 300 will carry out the following actions status register 2 12 alarm_volume_difference is activated event is recorded in th...

Страница 55: ......

Страница 56: ...8 2017 Rev A2 Wigersma Sikkema B V PO Box 109 NL 6980 AC Doesburg Leigraafseweg 4 NL 6983 BP Doesburg The Netherlands TEL 31 0 313 47 19 98 FAX 31 0 313 47 32 90 info wigersma sikkema com www wigersma...

Отзывы: