5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-25
Table 5.7 – I/A – AMO Default Settings
Parameter
Available
Settings
AMO Default Settings
IA1
IA2
IA3
Continuous
Irrigation
On, Off
Off
Off
Off
Pump
Selection
Venturi
Peristaltic
Peristaltic
Peristaltic
Peristaltic
Bottle Height
0–104 cm
0–42 inches
76 cm
30 inches
50 cm
20 inches
50 cm
20 inches
Max Vacuum
0–650
mmHg
500
mmHg
500
mmHg
15 mmHg
Max
Aspiration/
Mode
(Peristaltic)
0–60 cc/min
26 cc/min
Panel
26 cc/min
Linear
6 cc/min
Panel
Peristaltic
Pump Ramp
Threshold
10–100%
80%
80%
80%
Venturi Pump
On, Off
Off
Off
Off
Passive Reflux
(Fusion™
Fluidics Pack)
Yes, No
Yes
Yes
Yes
Passive Reflux
(Fusion™
Dual Pump
Pack)
Yes, No
Yes
Yes
Yes
Initial Mode/
Submode
Any of the
Submode
Names
Phaco 1
Phaco 1
Phaco 1
Fluidic Mode
Panel, Linear
Panel
Aspiration
Linear
Vacuum
Linear
Aspiration
Panel
Vacuum
Panel
Aspiration
Linear
Vacuum