5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-10
Table 5.1 – WHITESTAR
®
Technology Parameter Settings
Figure 5.9 – Duty Cycle
When the Variable WHITESTAR
®
Technology is used, different duty cycles are
applied as the footpedal moves through the power delivery zone. The zone is
divided into four equal size quadrants, and a different duty cycle can be applied in
each quadrant.
Setting
Pulse On
Time (ms)
Pulse Off
Time (ms)
Duty Cycle
Pulse Rate
(pps)
6/12
6
12
33%
55
4/8
4
8
33%
83
6/4
6
4
60%
100
6/8
6
8
43%
71
8/4
8
4
67%
83
4/24
4
24
14%
35
6/28
6
28
18%
29
6/24
6
24
20%
33
8/24
8
24
25%
31
6/18
6
18
25%
41