5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-18
Using Vitrectomy
In the left panel, press
VIT 1
or
VIT 2
to select a submode.
VIT 1
is selected by
default.
Note: Each time you select the vitrectomy mode you are prompted to prime the
vitrectomy handpiece. Refer to Chapter 4,
Equipment Operation
, Priming
for Vitrectomy for detailed information.
Figure 5.17 – Vitrectomy Mode Screen
1. In the main panel, press the
Up
or
Down
arrows to increase or decrease the
Aspiration Rate, Vacuum, or Cut Rate.
Note: If the Venturi pump is On, only the Vacuum and Cut Rate Settings are
shown.
2. To program the footpedal, go to
Configuration
,
Surgeons and Programs
,
Foot Pedal
,
Configuration, VIT
.
3. Select a
Submode
to program.
4. Edit the footpedal settings as needed.