Aquarius DIW Heaters Owner’s Manual
Ver. 1.0.0 | 29 Apr 2020 | P. 6
Subject to change without notice
7. Aquarius
®
systems with Recirculation Option is tested to ensure optimum functionality
and performance during recirculation.
a. The inputs and outputs of the pump and pump controller are checked and verified.
b. Communication between the pump, pump controller, and Aquarius
®
control system
are verified and checked for functionality.
c. The pump safeties and/or alarms are tested. This includes the Water Leaks Alarm,
which is standard on the Aquarius
®
system, to ensure that the supply of DIW to the
unit is shut off in the case that a critical leak condition is present.
d. The unit is allowed to operate with the controller “
Process Temperature Set Point
”
set to a value below the maximum temperature possible with the unit in Recirculation
to ensure optimal temperature performance during Recirculation.
e. All external signals to the Aquarius
®
system are tested.
f. Aquarius
®
pump air cooling systems are tested to ensure optimal pump performance.