Aquarius DIW Heaters Owner’s Manual
Ver. 1.0.0 | 29 Apr 2020 | P. 19
Subject to change without notice
11.4 Pump Air Cooling System Calibration
Calibration of the pump air cooling system is necessary once the air supply connection
for the pump air cooling system has been completed. This is achieved by regulating the
air pressure of the pump air cooling system when it is active. To perform this calibration
procedure access to the interior enclosure of the Aquarius
®
system with the system
powered ON is required. Please refer to the PF3000 controller manual for more information
on the HMI/Touch screen operation of the Aquarius
®
system.
Please take extreme cautionary measures as live voltage will be present in Aquarius
®
system during the Pump Air Cooling System Calibration procedure.
1. Take the appropriate steps necessary to have the Aquarius
®
system powered ON with the
enclosure door open. Please note that the Door Interlock switch will prevent the unit from
powering ON when the enclosure door is open.
2. Verify that an adequate steady supply of air is being supplied to the pump air cooling
system (60 psi maximum).
3. Locate the pump air cooling system pressure regulator/gauge in the heater compartment
of the Aquarius
®
system and take note of the air pressure reading on the pressure gauge.
Please refer to the Mechanical Layout Drawing in SECTION II for its specific location.
4. Turn the Aquarius
®
system ON, once again with the enclosure door open.
5. Verify that the Recirculation System is enabled which is indicated by an illuminated
yellow “Enabled” key on the System Status screen. If the Recirculation System is not
enabled press the DISABLED key.
6. Send the unit into Recirculation mode by providing a closed “Recirculation Signal”.
7. Once the unit is in Recirculation mode, activate the pump air cooling system temporarily
by pressing the “Pump Air Cooling System Calibration ON/OFF” key located on the
System Calibration screen.
8. The pump air cooling system will activate for approximately 60 seconds and then shut
off. Please note that the air pressure reading on the pressure gauge may drop when the
pump air cooling system is actively cooling.
9. With the pump air cooling system active, check the air pressure reading on the pressure
gauge and regulate the air pressure of the pump air cooling system to a minimum of 20
psi by adjusting the air pressure regulator accordingly.
Note: regulated air pressure of the pump air cooling system is dependent on the air supply
source pressure. Please provide a stable supply of air to the pump air cooling system.