background image

MICRO CENTRIFUGE
USER MANUAL

CODE 663-958

C

WESTLAB SMART

INSTRUCTION

MANUAL

1. Introduction

1

2. Intended Use

1

3. Features

1

4. Techinical Specifications

2

5. Standard Accessories

2

6. Safety Precautions

2

7. Installation

3

8. Standard Part Listing

4

9. Rotor Balancing

4

10. Interchageable Rotor

5

11. Changing the Rotor

5

12. Opening Centrifuge Lid

6

13. Operation

7

14. Care and Maintenance

10

15. Warranty Statement

10

*Please read this instructions

carefully before use.

Содержание 663-958C

Страница 1: ...nical Specifications 2 5 Standard Accessories 2 6 Safety Precautions 2 7 Installation 3 8 Standard Part Listing 4 9 Rotor Balancing 4 10 Interchageable Rotor 5 11 Changing the Rotor 5 12 Opening Centr...

Страница 2: ...e on lid opening for user safety Quick start stop for short spin Smooth and quiet running Tiny footprint It can squeeze easily on to a crowded counter top Interchangeable PCR rotor adaptor Low voltage...

Страница 3: ...4 C to 45 C Main supply voltage fluctuations not to exceed 10 of the nominal voltage Position at least 20 cm from walls 20 cm from ceiling and 20 cm between other products Make sure to always install...

Страница 4: ...ied micro centrifuge tubes made from plastic and designed to withstand centrifugal forces of at least 2000 g Do not operate the centrifuge without the rotor being properly attached to the shaft Re che...

Страница 5: ...e serious damage to the centrifuge 8 STANDARD PARTS LISTING ON OFF SWITCH NON SLIP RUBBER BASE POWER SOCKET PROTECTIVE LID ROTOR SAFETY SWITCH DISPLAY RELEASE BUTTON SPEED KNOB x 100RPM Min 60 TIMER K...

Страница 6: ...is located at the bottom of the rotor on the marked side To remove the rotor turn counter clockwise to loosen the screw and then gently pull the rotor up vertically Insert the new rotor from the top...

Страница 7: ...ter clockwise direction Trying to open the centrifuge lid by pulling the Lid Release button up is the wrong method of opening the lid Pulling the Lid Release button up will break the button or damage...

Страница 8: ...id accidents and damage to the centrifuge 4 Close the lid flip ON the front switch to start centrifugation 5 Once the centrifugation is completed turn the front switch to Off to gradually stop the rot...

Страница 9: ...nd rotor the actual RCF value will vary One can refer to the table mentioned in the section 7 to get the actual RCF of different rotors at certain given speed values 8 The unit is equipped with a 0 25...

Страница 10: ...ge to the centrifuge This centrifuge has an unique patented feature to protect the centrifuge its user the surroundings against such as accidental run When such an event occurs the sensor build inside...

Страница 11: ...us materials occur within the rotor or rotor chamber the unit should be disinfected This should be performed by qualified personnel with proper protective equipment Do not use any cleaning or decontam...

Страница 12: ...turned for service should be sent with postage prepaid in the original packaging or other suitable packaging padded to avoid damage We will not be responsible for damage incurred by improper packaging...

Отзывы: