9
When you now release the "CAL" key the scale will display "C XXX" (XXX is the weight of the
standard weight. Example: display "C 200" should be put on the standard weight of 200g), put
on the calibration weights, after the weight of the calibration weight displayed and the stable
weight symbol "g" shows that means the calibration is complete. If press the "CAL" key shows
"C----F", the zero point is unstable, and the "TARE" key can be re pressed to show the return to
zero. Now press the "CAL" key again.
Adjustment with recommended adjustment weight (default setting) Weight value of the
required adjustment weight seeschap. 1 “Technical data”. Weights of different nominal values
may be used for adjustment but are not optimal for technical measuring.
Operating elements
1. Weighing pan 2. Pan support ring
3. Wind protection 4. Underfloor weighing widget
5. Sealing cap 6. Adjustable feet 7. RS232 Interface 8. Level Indicator
9. Power socket 10. Power Cord 11. ON/OFF key 12.Calibration key
13.Unit change key 14.Counting key 15.Print key 16.Tare key
6
Basic Operation
After new purchase of the balance or after a long storage, you have to calibrate the balance
before the first commissioning. Level the balance with the adjustable feet horizontal until the air
bubble in the level is in the center of the Level Indicator. Give the balance after changing the
location a warm-up time min. 1 hour before using. This time is necessary for the balance to
adapt to the eventual temperature.
6.1
[Power On/OFF]
Press the[On/OFF]button to turn on the balance. The display shows the message "
ON
". Then the
max. weighing range is displayed. The balance will carry out a self-test (
F----1
to
F----9
). When
the "0.0000 g" in display appears, the balance is ready for operation.
Press the [On/OFF]button and keep it pressed, until the display shows the message “
OFF
”. The
balance will switch off then.