When you now release the "CAL" key the scale will display "C XXX" (XXX is the weight of the 

standard weight. Example: display "C 200" should be put on the standard weight of 200g), put 

on the calibration weights, after the weight of the calibration weight displayed and the stable 

weight symbol "g" shows that means the calibration is complete. If press the "CAL" key shows 

"C----F", the zero point is unstable, and the "TARE" key can be re pressed to show the return to 

zero. Now press the "CAL" key again. 

 

Adjustment with recommended adjustment weight (default setting) Weight value of the 

required adjustment weight seeschap. 1 “Technical data”. Weights of different nominal values 

may be used for adjustment but are not optimal for technical measuring.  

Operating elements 

 

1. Weighing pan  2. Pan support ring 

3. Wind protection  4. Underfloor weighing widget 

5. Sealing cap               6. Adjustable feet        7. RS232 Interface  8. Level Indicator 

9. Power socket           10. Power Cord           11. ON/OFF key      12.Calibration key  

13.Unit change key    14.Counting key          15.Print key              16.Tare key 

6

 

Basic Operation 

After new purchase of the balance or after a long storage, you have to calibrate the balance 

before the first commissioning. Level the balance with the adjustable feet horizontal until the air 

bubble in the level is in the center of the Level Indicator. Give the balance after changing the 

location a warm-up time min. 1 hour before using. This time is necessary for the balance to 

adapt to the eventual temperature. 

6.1

 

[Power On/OFF] 

Press the[On/OFF]button to turn on the balance. The display shows the message "

ON

". Then the  

max. weighing range is displayed. The balance will carry out a self-test (

F----1

 to 

F----9

). When 

the "0.0000 g" in display appears, the balance is ready for operation. 

Press the [On/OFF]button and keep it pressed, until the display shows the message “

OFF

”. The 

balance will switch off then. 

Содержание MAB 201

Страница 1: ...1 ANALYTICAL BALANCE MAB 201 220g 0 1mg MAB 301 320g 0 1mg...

Страница 2: ...l Commissioning 8 5 6 Adjustment 8 5 7 Calibration 8 5 7 1 Internal calibration 8 5 7 2 Internal calibration 8 6 Basic Operation 9 6 1 Power On OFF 9 6 2 TARE 10 6 3 CAL calibration with external weig...

Страница 3: ...cription of the data transfer 13 9 3 Remote control instructions 13 9 4 RS232 Pinout 9 pin D Sub 14 10 Configuration Explain of the function setting 14 11 Maintenance and Servicing 16 11 1 Cleaning 16...

Страница 4: ...s very wide range of functions weighing parts counting or density determination Its a built in standard weight and a precision automatic loading device our computer controlled mechanism with precise m...

Страница 5: ...32 interface enables you to automatically send the weighing data to PC Powered by the external 3 pin power cord 2 1 Proper Use The balance you purchased is intended to determine the weighing value of...

Страница 6: ...balance is a precision measuring instrument and must be handled with great care Violation of limit of weighing range or impact can cause permanent damage to the balance The balance needs a warm up ti...

Страница 7: ...ue to open windows and doors Avoid jarring during weighing Protect the balance against high humidity vapors and dust Do not expose the device to extreme dampness for longer periods of time Non permitt...

Страница 8: ...ce must be coordinated in compliance with the underlying physical weighing principle to the existing acceleration due to gravity at its place of location only if the balance has not already been adjus...

Страница 9: ...ghing widget 5 Sealing cap 6 Adjustable feet 7 RS232 Interface 8 Level Indicator 9 Power socket 10 Power Cord 11 ON OFF key 12 Calibration key 13 Unit change key 14 Counting key 15 Print key 16 Tare k...

Страница 10: ...unction button 6 4 1 Weighing unit selection weighing mode Under the weighing mode use the UNIT key to choose weighing units g gram ct carat ozt troy ounce oz ounce GN gran dwt pennyweight and lb poun...

Страница 11: ...hat might be damaged underneath the load After completing the underfloor weighing the opening on the bottom of the balance must always be closed dust protection 7 Density determination Hydrostatic wei...

Страница 12: ...artially immersed in a fluid is buoyed up by a force equal to the weight of the fluid displaced by the object The volume V of the glass sinker is same as the volume of the displaced liquid Buoyant for...

Страница 13: ...4800 and9600bit s default setting is 600 Miniature plug is necessary 9 pin D Sub 9 2 Description of the data transfer Symbols minus sign for positive numbers this is output as blank Data value of wei...

Страница 14: ...er terminating equipment Balance Pin Typical purpose DE 9 DTE e g PC Carries data from DCE to DTE 2 Input Carries data from DTE to DCE 3 Output 5 27 default setting the codes shown in VB as following...

Страница 15: ...ger The number corresponds to the first data signal to the command control of the computer Default setting is 27 1Bhex e C5 setting the mode for the backlight 0 always on 1 automatic after the weight...

Страница 16: ...nnection has been interrupted mains cable not plugged in faulty Power supply is interrupted The displayed weight is constantly changing Draught air movement Table floor vibrations Weighing plate has c...

Отзывы: