2.3

 

Control of inspection, measuring and test equipment 

As part of quality assurance the measuring-related properties of the balance and the available 

adjusting weight must be checked regularly. The responsible user must define a suitable interval 

as well as the nature and scope of this test.  

3

 

Precautions 

 

Environmental conditions

: only use indoor, check the detailed environment conditions for 

temperature and humidity in the technical details above. 

 

Do not use the scale in any high temperature fluctuations and avoid strong airflows, vibrations 

and dust. The balance could never get wet and make sure that the balance is not used in an 

environment with extreme temperature and humidity. 

 

Do not use the balance in explosive conditions. 

 

Only use the extender cable, which with ground wire. 

 

If the power line cable is damaged, the appliance must be unplugged immediately and the 

power line cable be replaced. 

 

Place the balance on a stable surface and set it horizontally with the help of the adjustable feet 

and the spirit level. 

 

The electronic balance is a precision measuring instrument and must be handled with great care. 

Violation of limit of weighing range or impact can cause permanent damage to the balance. 

 

The balance needs a warm up time for about 1 hour before using. 

 

Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened 

with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft 

cloth. 

 

Do not overload the balance more than 20% of maximum load (Max). Do not press the pan with 

a hand! 

 

By maintenance please notice the information in 

Chapter “Maintenance and Servicing”

 

If the balance seems not to operate properly, unplug it from the mains and do not use until 

checked by authorized service. 

 

All repairs and necessary regulations can be made by authorized personnel only. 

4

 

Transportation and Storage 

4.1

 

Acceptance check 

When receiving the appliance, please check packaging immediately, and the appliance itself 

when unpacking for possible visible damage. 

4.2

 

Packing 

 

Keep all parts of the original packaging for a possibly required return 

 

Only use original packaging for returning. 

 

Prior to dispatch disconnect all cables and remove loose/mobile parts. 

 

Reattach possibly supplied transport securing devices. 

Содержание MAB 201

Страница 1: ...1 ANALYTICAL BALANCE MAB 201 220g 0 1mg MAB 301 320g 0 1mg...

Страница 2: ...l Commissioning 8 5 6 Adjustment 8 5 7 Calibration 8 5 7 1 Internal calibration 8 5 7 2 Internal calibration 8 6 Basic Operation 9 6 1 Power On OFF 9 6 2 TARE 10 6 3 CAL calibration with external weig...

Страница 3: ...cription of the data transfer 13 9 3 Remote control instructions 13 9 4 RS232 Pinout 9 pin D Sub 14 10 Configuration Explain of the function setting 14 11 Maintenance and Servicing 16 11 1 Cleaning 16...

Страница 4: ...s very wide range of functions weighing parts counting or density determination Its a built in standard weight and a precision automatic loading device our computer controlled mechanism with precise m...

Страница 5: ...32 interface enables you to automatically send the weighing data to PC Powered by the external 3 pin power cord 2 1 Proper Use The balance you purchased is intended to determine the weighing value of...

Страница 6: ...balance is a precision measuring instrument and must be handled with great care Violation of limit of weighing range or impact can cause permanent damage to the balance The balance needs a warm up ti...

Страница 7: ...ue to open windows and doors Avoid jarring during weighing Protect the balance against high humidity vapors and dust Do not expose the device to extreme dampness for longer periods of time Non permitt...

Страница 8: ...ce must be coordinated in compliance with the underlying physical weighing principle to the existing acceleration due to gravity at its place of location only if the balance has not already been adjus...

Страница 9: ...ghing widget 5 Sealing cap 6 Adjustable feet 7 RS232 Interface 8 Level Indicator 9 Power socket 10 Power Cord 11 ON OFF key 12 Calibration key 13 Unit change key 14 Counting key 15 Print key 16 Tare k...

Страница 10: ...unction button 6 4 1 Weighing unit selection weighing mode Under the weighing mode use the UNIT key to choose weighing units g gram ct carat ozt troy ounce oz ounce GN gran dwt pennyweight and lb poun...

Страница 11: ...hat might be damaged underneath the load After completing the underfloor weighing the opening on the bottom of the balance must always be closed dust protection 7 Density determination Hydrostatic wei...

Страница 12: ...artially immersed in a fluid is buoyed up by a force equal to the weight of the fluid displaced by the object The volume V of the glass sinker is same as the volume of the displaced liquid Buoyant for...

Страница 13: ...4800 and9600bit s default setting is 600 Miniature plug is necessary 9 pin D Sub 9 2 Description of the data transfer Symbols minus sign for positive numbers this is output as blank Data value of wei...

Страница 14: ...er terminating equipment Balance Pin Typical purpose DE 9 DTE e g PC Carries data from DCE to DTE 2 Input Carries data from DTE to DCE 3 Output 5 27 default setting the codes shown in VB as following...

Страница 15: ...ger The number corresponds to the first data signal to the command control of the computer Default setting is 27 1Bhex e C5 setting the mode for the backlight 0 always on 1 automatic after the weight...

Страница 16: ...nnection has been interrupted mains cable not plugged in faulty Power supply is interrupted The displayed weight is constantly changing Draught air movement Table floor vibrations Weighing plate has c...

Отзывы: