Maintenance and Servicing

 

11.1

 

Cleaning 

Before cleaning, please disconnect the appliance from the operating voltage. 

Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth dampened 

with mild soap suds. Ensure that no liquid penetrates into the device and wipe with a dry soft 

cloth. 

Loose residue sample/powder can be removed carefully with a brush or manual vacuum cleaner.  

Remove spillages immediately. 

11.2

 

Maintenance and repair 

The appliance may only be opened by trained service technicians who are authorized by us 

Before opening, disconnect from power supply. 

12

 

Trouble Shooting 

Failure 

Possible Cause 

The display does 
not light up. 

The balance is not switched on. 
The mains supply connection has been interrupted (mains cable not plugged 
in/faulty). 
Power supply is interrupted. 

The displayed 
weight is constantly 
changing 

Draught /air movement. 
Table / floor vibrations. 
Weighing plate has contact with other objects.  

Electromagnetic fields / static charging (choose different location/switch off 
interfering device if possible). 

The weighing result 
is obviously wrong 

The display of the balance is not at zero. 
Adjustment is no longer correct. 
Great temperature fluctuations. 
Warm-up time was ignored. 

Electromagnetic fields / static charging (choose different location/switch off 
interfering device if possible). 

F----L 

Lower zero position of sensor signal (re-Adjustment). 

F----H 

Weighing capacity is exceeded. 

C----L 

Lower zero position of sensor signal (re-Adjustment). 

C----F 

Calibration error, instability or not at zero point during calibrating. 

C----H 

Calibration error, items on the pan or at higher zero point during calibrating. 

 

****************** 

 
 
 
 
 

 
 
 
 
 

14092020 

Содержание MAB 201

Страница 1: ...1 ANALYTICAL BALANCE MAB 201 220g 0 1mg MAB 301 320g 0 1mg...

Страница 2: ...l Commissioning 8 5 6 Adjustment 8 5 7 Calibration 8 5 7 1 Internal calibration 8 5 7 2 Internal calibration 8 6 Basic Operation 9 6 1 Power On OFF 9 6 2 TARE 10 6 3 CAL calibration with external weig...

Страница 3: ...cription of the data transfer 13 9 3 Remote control instructions 13 9 4 RS232 Pinout 9 pin D Sub 14 10 Configuration Explain of the function setting 14 11 Maintenance and Servicing 16 11 1 Cleaning 16...

Страница 4: ...s very wide range of functions weighing parts counting or density determination Its a built in standard weight and a precision automatic loading device our computer controlled mechanism with precise m...

Страница 5: ...32 interface enables you to automatically send the weighing data to PC Powered by the external 3 pin power cord 2 1 Proper Use The balance you purchased is intended to determine the weighing value of...

Страница 6: ...balance is a precision measuring instrument and must be handled with great care Violation of limit of weighing range or impact can cause permanent damage to the balance The balance needs a warm up ti...

Страница 7: ...ue to open windows and doors Avoid jarring during weighing Protect the balance against high humidity vapors and dust Do not expose the device to extreme dampness for longer periods of time Non permitt...

Страница 8: ...ce must be coordinated in compliance with the underlying physical weighing principle to the existing acceleration due to gravity at its place of location only if the balance has not already been adjus...

Страница 9: ...ghing widget 5 Sealing cap 6 Adjustable feet 7 RS232 Interface 8 Level Indicator 9 Power socket 10 Power Cord 11 ON OFF key 12 Calibration key 13 Unit change key 14 Counting key 15 Print key 16 Tare k...

Страница 10: ...unction button 6 4 1 Weighing unit selection weighing mode Under the weighing mode use the UNIT key to choose weighing units g gram ct carat ozt troy ounce oz ounce GN gran dwt pennyweight and lb poun...

Страница 11: ...hat might be damaged underneath the load After completing the underfloor weighing the opening on the bottom of the balance must always be closed dust protection 7 Density determination Hydrostatic wei...

Страница 12: ...artially immersed in a fluid is buoyed up by a force equal to the weight of the fluid displaced by the object The volume V of the glass sinker is same as the volume of the displaced liquid Buoyant for...

Страница 13: ...4800 and9600bit s default setting is 600 Miniature plug is necessary 9 pin D Sub 9 2 Description of the data transfer Symbols minus sign for positive numbers this is output as blank Data value of wei...

Страница 14: ...er terminating equipment Balance Pin Typical purpose DE 9 DTE e g PC Carries data from DCE to DTE 2 Input Carries data from DTE to DCE 3 Output 5 27 default setting the codes shown in VB as following...

Страница 15: ...ger The number corresponds to the first data signal to the command control of the computer Default setting is 27 1Bhex e C5 setting the mode for the backlight 0 always on 1 automatic after the weight...

Страница 16: ...nnection has been interrupted mains cable not plugged in faulty Power supply is interrupted The displayed weight is constantly changing Draught air movement Table floor vibrations Weighing plate has c...

Отзывы: