5.3

 

Mains connection 

Before connecting the mains power cord check if the printed voltage value is the same as the 

local supply voltage. 

Only use the original supplied mains power cord. Otherwise there is a risk that the Scale be 

impaired. 

5.4

 

Connection of peripheral devices 

The balance may send data to a computer or a printer through RS232C port. 

When cooperating with the balance, a computer should be equipped with a program which 

enables processing data from the balance. 

Before connecting or disconnecting of additional devices (printer, PC) to the data interface, 

always disconnect the balance from the power supply. 

5.5

 

Initial Commissioning 

In order to obtain exact results with the electronic balances, your balance must have reached 

the operating temperature (warming up time 1 hour). During this warming up time the balance 

must be connected to the power supply 

The accuracy of the balance depends on the local acceleration of gravity. 

Please read the information carefully in chap. “Adjustment”. 

5.6

 

Adjustment 

As the acceleration value due to gravity is not the same at every location on earth, each balance 

must be coordinated - in compliance with the underlying physical weighing principle - to the 

existing acceleration due to gravity at its place of location (only if the balance has not already 

been adjusted to the location in the factory). This adjustment process must be carried out for 

the first commissioning, after each change of location as well as in case of fluctuating 

environment temperature. To receive accurate measuring values it is also recommended to 

adjust the balance periodically in weighing operation.  

5.7

 

Calibration 

5.7.1

 

Internal calibration: 

When the balance display is “0” then press the “CAL” key.  

The balance turns into the automatic internal calibration status and displays “CAL”, now the 

internal weight loading and unloading device for automatic calibration will start their work. 

After 30 seconds the display “0”, that means the calibration is finished. In the normal use, the 

internal calibration should be used to calibrate the balance. 

 

When there was no calibration within the last 3 hours the scale will calibrate itself. 

5.7.2

 

Internal calibration: 

External Calibration: 

First you should get ready the standard weight for external calibration in the required OIML-

Tolerance (E2 for example). When the balance display is “0” press and hold the “CAL” key until 

the display changes from "CAL-INT" to "CAL-E". 

Содержание MAB 201

Страница 1: ...1 ANALYTICAL BALANCE MAB 201 220g 0 1mg MAB 301 320g 0 1mg...

Страница 2: ...l Commissioning 8 5 6 Adjustment 8 5 7 Calibration 8 5 7 1 Internal calibration 8 5 7 2 Internal calibration 8 6 Basic Operation 9 6 1 Power On OFF 9 6 2 TARE 10 6 3 CAL calibration with external weig...

Страница 3: ...cription of the data transfer 13 9 3 Remote control instructions 13 9 4 RS232 Pinout 9 pin D Sub 14 10 Configuration Explain of the function setting 14 11 Maintenance and Servicing 16 11 1 Cleaning 16...

Страница 4: ...s very wide range of functions weighing parts counting or density determination Its a built in standard weight and a precision automatic loading device our computer controlled mechanism with precise m...

Страница 5: ...32 interface enables you to automatically send the weighing data to PC Powered by the external 3 pin power cord 2 1 Proper Use The balance you purchased is intended to determine the weighing value of...

Страница 6: ...balance is a precision measuring instrument and must be handled with great care Violation of limit of weighing range or impact can cause permanent damage to the balance The balance needs a warm up ti...

Страница 7: ...ue to open windows and doors Avoid jarring during weighing Protect the balance against high humidity vapors and dust Do not expose the device to extreme dampness for longer periods of time Non permitt...

Страница 8: ...ce must be coordinated in compliance with the underlying physical weighing principle to the existing acceleration due to gravity at its place of location only if the balance has not already been adjus...

Страница 9: ...ghing widget 5 Sealing cap 6 Adjustable feet 7 RS232 Interface 8 Level Indicator 9 Power socket 10 Power Cord 11 ON OFF key 12 Calibration key 13 Unit change key 14 Counting key 15 Print key 16 Tare k...

Страница 10: ...unction button 6 4 1 Weighing unit selection weighing mode Under the weighing mode use the UNIT key to choose weighing units g gram ct carat ozt troy ounce oz ounce GN gran dwt pennyweight and lb poun...

Страница 11: ...hat might be damaged underneath the load After completing the underfloor weighing the opening on the bottom of the balance must always be closed dust protection 7 Density determination Hydrostatic wei...

Страница 12: ...artially immersed in a fluid is buoyed up by a force equal to the weight of the fluid displaced by the object The volume V of the glass sinker is same as the volume of the displaced liquid Buoyant for...

Страница 13: ...4800 and9600bit s default setting is 600 Miniature plug is necessary 9 pin D Sub 9 2 Description of the data transfer Symbols minus sign for positive numbers this is output as blank Data value of wei...

Страница 14: ...er terminating equipment Balance Pin Typical purpose DE 9 DTE e g PC Carries data from DCE to DTE 2 Input Carries data from DTE to DCE 3 Output 5 27 default setting the codes shown in VB as following...

Страница 15: ...ger The number corresponds to the first data signal to the command control of the computer Default setting is 27 1Bhex e C5 setting the mode for the backlight 0 always on 1 automatic after the weight...

Страница 16: ...nnection has been interrupted mains cable not plugged in faulty Power supply is interrupted The displayed weight is constantly changing Draught air movement Table floor vibrations Weighing plate has c...

Отзывы: