Safety instructions and warnings
14/90
Environmental conditions
Optimum ambient temperature:
between 59 °F (15 °C) and ~ 90°F (~ 32°C)
Storage temperature: Between -15 °C (5 ° F) and +60 °C (140 ° F)
(the well, pump and the hoses must not contain any water)
Relative humidity: Between 30 % and 75 %
Air pressure: between 794 hPa and 1060 hPa
Water and cooling conditions
Permissible water pressure (water inlet): 116 PSI / 8 bar
Maximum coolant temperature: 16°C (61 °F)
The device is not to be used outdoors.
The device is not suitable for operation in mobile facilities (ships,
buses, trains or oil platform). The unit has to be adapted in order to
guarantee trouble-free operation when used in mobile facilities.
Adaptation of the unit is required for operating altitudes greater than
6500 ft / 2000 m above sea level in order to guarantee trouble-free op-
eration.
Please contact the customer service department before putting the
device into operation - see page 2.
2.6.5
Commissioning
Initial commissioning will be undertaken by the manufacturer's own
service personnel or by an authorized service company. The device
will be handed over ready for use.
If the device has not been used for a longer period of time then it must
first be inspected by our customer service staff or by another author-
ized company before being put back into operation again.
The device is only to be used if it is working correctly!
None of the safety devices (e.g. switch) or safety warnings are to be
removed or deactivated as this would impair the safe operation of the
device!
The unit may only be put into operation if the well has been filled with
the specified amount of water.
The unit may not be connected to the power supply during filling!
The unit may no longer be moved once the well has been filled!
A water cooling system must be connected to the device if it is to be
used for more than 1 hour per day at maximum pressure. The pres-
sure hoses must not be removed. Please contact the customer ser-
vice department without fail before putting the device into operation -
see page 2.
The device may only be operated with the pressure hose included in
the scope of delivery.
A water shut-off valve must be used.
High frequency mobile communication systems can interfere with the
way the device functions.
Heat therapy devices (microwave therapy devices, etc.) installed in the
vicinity may lead to electromagnetic interference. Contact our cus-
tomer service department if this happens – see page 2.