Safety instructions and warnings
06/2021
1026388-00A
Operating instructions – 9/82
1.2
Use
1.2.1
Purpose
wellsystem MEDWAVE TOUCH permits a dry water massage through the
physical stimulation of certain massage zones (groups of muscles) of the
human body that have been chosen according to physiological correctness
using controlled and temperature-controlled water pressure via an electric
mat.
The medical device is indicated for therapy or to assist in the therapy of mus-
cular / musculoskeletal disorders and dysfunctions and the pain caused by
this.
The therapy in the form of a dry water massage can be carried out on adult
patients of all body sizes, ages and constitution. Infants and small children up
to the age of 7 may not be subject to a therapy using the wellsystem medical
products. Children and adolescents from 8 to 17 years of age are only al-
lowed to use wellsystem medical products in agreement with a parent or
guardian and after consulting a doctor.
The therapy with wellsystem MEDWAVE TOUCH must be carried out and
accompanied by trained specialised staff. The medical product is not indicat-
ed for use in private households.
The duration and type of treatment must be determined by your doctor.
The unit may only be used by people with a heart pacemaker after consulta-
tion with a doctor.
1.2.2
Intended use
This device is used to apply a dry water massage to one adult patient using
the programmes described in these operating instructions. More detailed
information can be found on pages 26 and 30.
The device is only intended for commercial use (e.g. for use in doctors surger-
ies, by physical therapists or in hospitals and clinics) and not for domestic
use.
The unit may only be operated by trained specialist personnel. Patients may
not operate the unit.
A water cooling system must be connected to the device if it is to be used for
more than 1 hour per day at maximum pressure. The pressure hoses must
not be removed. Please contact the customer service department without fail
before putting the device into operation - see page 2.
The device may only be operated with the pressure hose included in the
scope of delivery.
Any other use shall be considered improper. The manufacturer cannot be
held liable for damage or injuries resulting from this. The operator bears the
sole risk for this.
Proper use also includes compliance with the manufacturer’s instructions and
its operating and maintenance conditions. The device may only be operated,
maintained and repaired by trained professionals familiar with this device and
who have been informed of the dangers involved.